Detailed instructions for use are in the User's Guide.
Instruction Manual
lpha pH 190
1/8 DIN pH / ORP Controller with Temperature display and Transmitter
Technology Made Easy ...
68X276106 Rev 0 11/2004
Preface This manual serves to explain the use of the lpha pH 190 controller/transmitter. This manual functions in two ways: first, as a step-by-step guide to help you operate the meter; second, it serves as a handy reference guide. This manual is written to cover as many anticipated applications of the lpha pH 190 controller/transmitter as possible. If there are doubts in the use of the lpha pH 190 controller/transmitter, do not hesitate to contact the nearest Eutech Instruments Authorized Distributor. Eutech Instruments cannot accept any responsibility for damage or malfunction to the controller/transmitter caused by improper use of the instrument. Remember to fill in the guarantee card and mail it to your Authorized Distributor or Eutech Instruments Pte Ltd. The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on the part of Eutech Instruments Pte Ltd. Copyright© Nov 2004 Eutech Instruments Pte Ltd. All rights reserved.
Rev 0 11/2004
TABLE OF CONTENTS
1 2 3 3.1 3.2 3.3 3.4 4 4.1 4.2 5 6 7 8 9 9.1 9.2 9.3 9.4 9.5 9.6 9.7 9.8 9.9 10 11 12 13 14 15 INTRODUCTION ........................................................................................... 1 SAFETY INFORMATION .............................................................................. 2 OVERVIEW.................................................................................................... 3 FRONT PANEL....................................................................................... 3 BACK PANEL......................................................................................... 4 WIRING................................................................................................. 5 PANEL-MOUNTING THE CONTROLLER ..................................................... 6 MEASUREMENT MODE............................................................................... 7 MEASUREMENT MODE DISPLAY.............................................................. 7 SET POINTS ADJUSTMENTS................................................................... 7 PASSWORD.................................................................................................. 9 PH CALIBRATION ...................................................................................... 11 ORP CALIBRATION ................................................................................... 14 TEMPERATURE CALIBRATION................................................................ 15 SETUP MODE ............................................................................................. 16 GENERAL INFORMATION...................................................................... 16 SETUP MODE OVERVIEW ..................................................................... 17 SET POINT 1 Â P1.0............................................................................ 18 SET POINT 2 Â P2.0............................................................................ 20 CONFIGURATION Â P3.0...................................................................... 22 VIEWING CALIBRATION POINTS: PH OR ORP Â P4.0 ............................ 24 VIEWING PH ELECTRODE DATA Â P5.0................................................ 25 VIEWING ORP ELECTRODE DATA Â P5.0 ............................................ 26 CONTROLLER RESET Â P6.0 ............................................................... 27 RELAYS....................................................................................................... 28 TRANSMITTER FUNCTION ....................................................................... 28 ERROR MESSAGES................................................................................... 28 SPECIFICATIONS....................................................................................... 29 ACCESSORIES........................................................................................... 30 GENERAL INFORMATION......................................................................... 35
1 INTRODUCTION
Thank you for purchasing the lpha pH 190 DIN pH/ORP Controller. This controller is part of a series of quality process controllers available from Eutech Instruments. These sturdy, economical pH/ORP controllers are designed with the features and reliability of a much more expensive instrument. Your controller includes:
· ·
Removable terminal blocks for easy connections; Two mounting brackets for easy panel mounting;
When shipped, the controller is set to the pH mode. See Setup program P3.1 for directions on selecting ORP mode. Some features of this controller are: Two set points, two SPDT relay operation Scrolling, 14-segment LED guides user easily through setup functions Reliable power supply from 85 to 260 V AC, 50/60 Hz or DC withstands voltage fluctuations · Push-button operation from the front panel · Two-point pH calibration, offset temperature and ORP calibration · Adjustable hysteresis band prevents rapid contact switching near set-point · Selectable automatic or manual temperature compensation · Two-level password protection · Removable terminal strips for quick and easy connections · Built-in non volatile memory retains setup even if power fails, and lets you configure unit before installation · Isolated 4-20 mA output for remote monitoring or hard copy recording
· · ·
-1-
2 SAFETY INFORMATION
The Eutech Controller/Transmitter shall be installed and operated only in the manner specified in the Instruction manual. Only skilled, trained or authorized person should carry out installation, setup and operation of the instrument. Before powering up the unit, make sure that power source it is connected to, is as specified in the top label. Failure to do so may result in a permanent damage to the unit. The unit has live and exposed parts inside. If it has to be opened, make sure that the power to the unit is off and disconnected. The unit is Fuse protected. In the event the fuse has to be replaced, use only those as specified in the manual. The degree of protection against electric shock will be achieved only by observance of the corresponding installation rules.
-2-
3 OVERVIEW 3.1 Front Panel
The front panel consists of a 4-digit LED display, 8 LED annunciators and 4 keys. Annunciators 1. REL 1 2. REL 2 3. MEAS 4. CAL 5. SETUP 6. pH 7. oC 8. mV Keys 9. MODE 10. CAL / ENTER 11. SP1 / SP2 / Use to toggle between Measurement modes (pH/ORP and temperature). Use to enter Calibration mode. Use also to enter into levels of the Setup mode and to confirm changes made. Use to enter Set Point 1 (SP1) or Set Point 2 (SP2) adjustment mode. Use as increment or decrement keys during Calibration and Setup modes. Use both keys together in Calibration and Setup modes to escape to Measurement mode. Displayed when Relay 1 is activated. Displayed when Relay 2 is activated. Displayed in measurement mode. Displayed in calibration mode. Displayed in setup mode. Unit of the displayed parameter (pH). Unit of the displayed parameter (temperature). Unit of the displayed parameter (ORP).
12.14-segment display
SP1
CAL ENTER
SP2
MODE
pH 190 Series
-3-
3.2 Back Panel
The back panel consists of four different connectors that can be used with removable terminal blocks (included):
19
18
17
16
15
14
13 J3
D
pH/ORP
Pt 100 - IN Pt 100 - GND NC NC NC NC Pt 100 - SENSE
C
WARNING
12
11
10
9
8
7
6
5
4
3
2
1 J1
B
J2
RELAY 1 - POLE RELAY 2 - NO RELAY 1 - NO RELAY 1 - NC RELAY 2 - POLE RELAY 2 - NC GROUND N L LIQUID GND
A
A. 3 pin connector (for power supply) B. 9 pin connector (for relays) C. 7 pin connector (for temperature sensor) D. BNC connector (for pH or ORP electrode) A. 1. L- Live 2. N- Neutral 3. E- Earth Ground C. 13. Pt 100- SENSE 14. Pt 100- IN 15. Pt 100-GND 16. NC 17. NC 18. NC 19. NC BNC for pH/ORP
+ 4 - 20 mA
- 4 - 20 mA
B.
4. Liquid GND 5. RELAY 2- NC 6. RELAY 2- POLE 7. RELAY 2- NO 8. RELAY 1- NC 9. RELAY 1- POLE 10. RELAY 1- NO 11. Â 4 to 20 mA 12. +4 to 20 mA
-4-
D.
3.3 Wiring
Caution: Ensure electrical mains are disconnected before proceeding.
· · ·
1. Connect the power supply to the three-pin terminal block (A) VAC protective ground wire = 3 VAC neutral wire = 2 VAC live wire = 1 The lpha pH 190 controller accepts voltages from 85 to 260 VAC, 50/60 Hz or DC. 2. Connect the Pt 100 leads to terminals 13 and 15 (terminal block C). Either wire can be connected to either terminal. Terminals 13 and 14 must be shunted unless using a 3-wire RTD. NOTE: pH 190 is factory set for manual temperature compensation. ATC can be selected in Setup mode of Program P3.3. 3. Slide the BNC connector of the pH (ORP) probe to the BNC connector on the back of the controller. Turn the notches of the connector until they lock into place. 4. Power on the controller. The display automatically shows the pH (ORP) reading, and the pH (ORP) annunciator lights. NOTE: In the event Pt 100 is not connected or the connection is broken in the ATC mode, the pH/ORP display flashes to alert you.
-5-
3.4 Panel-mounting the controller
The supplied mounting hardware allows surface mounting to all panels and protective enclosures. Mounting cut-out size is 91 x 45 mm. To attach the mounting to the controller: 1. Align the catch to the side of the controller, and insert threaded rods through catch.
BACK VIEW
2. Screw the threaded-rod through the catch in a clock-wise direction. Tighten until the catch holds the controller firmly against the back of the panel or protective housing. Repeat on the other side.
Wall pa Threaded Catch
TOP VIEW
SIDE VIEW
-6-
Threaded rod
Catch
approx. 100 approx. 154.20
4 MEASUREMENT MODE
The lpha pH 190 controller is capable of taking pH (or ORP) measurement with Automatic (ATC) or Manual (MTC) Temperature Compensation. The measurements are displayed distinguishingly by the annunciators on the front panel.
4.1 Measurement mode display
Press the MODE key to toggle between pH (or ORP) and Temperature measurement mode. pH (ORP) measurement mode The controller starts up in the mode when last powered off (pH or ORP).The pH (or ORP) annunciator will light up. For selection of pH or ORP measurement mode, see Setup program P3.1. Temperature Measurement mode Press MODE key once to view the temperature measurement. The °C annunciator lights up. The display shows ATC or MTC, then the current measured temperature. For selection of temperature compensation ATC or MTC, see Setup program P3.3. NOTE: After pressing the MODE key to display Temperature, if there is no further key-press, the controller will automatically revert to pH (ORP) measurement mode after about 30 seconds.
4.2 Set Points Adjustments
You can make quick set points adjustments with the direct access of the Set Points adjustment modes (SP1 and/or SP2). By just pressing the SP1/ or SP2/key, you can enter the Set Point adjustment mode and set a new pH (or ORP) value that will cause your controller to activate. Set Point 1 (SP1) adjustment mode This lets you adjust the pH (or ORP) value in Set Point 1. If this value is crossed, the set point relay 1 LED will light. 1. Press the SP1/ key. The screen will scroll P1.1, SP1, and then the current set point value. 2. Press the or keys and adjust first relay set point. You can adjust it in 0.01 increments from 0.00 to 14.00 pH (-999 to 1000 mV). 3. Press ENTER to confirm and return to the measurement mode. NOTE: Press and keys together (ESCAPE) at anytime, to return to Measurement mode.
-7-
Set Point 2 (SP2) adjustment mode This lets you adjust the pH (or ORP) value in Set Point 2. If this value is crossed, the set point relay 2 LED will light. 1. 2. 3. Press the SP2/ key. The screen will scroll P2.1, SP2, and then the current set point value. Press the or keys and adjust second relay set point. You can adjust it in 0.01 increments from 0.00 to 14.00 pH (-999 to 1000 mV). Press ENTER to confirm and return to the measurement mode.
NOTE: Press and keys together (ESCAPE) at anytime, to return to Measurement mode. NOTE: These modes are only for adjusting relay 1 and/or relay 2 set points values. To set the relays as low or high set points or to set its hysteresis values, you have to make the adjustments from the Setup mode as in Section 9.3 and Section 9.4.
-8-
5 PASSWORD
The Setup mode is to be accessed by entering a password code. The direct accessed Set Point adjustment mode (SP1 and SP2) and the Calibration mode can also be accessed through this password code procedure. The lpha pH 190 controller features two separate passwords: · pH (or ORP) calibration mode and temperature offset password = 011 · Setup program password = 022 To enter the password: 1. Press the MODE key, and within 1 second, press the CAL/ENTER key. The d ...