Detailed instructions for use are in the User's Guide.
Instruction Manual
pH/Ion 510
Bench pH/Ion/mV Meter
68X090811 Rev. 6 08/04
Technology Made Easy ...
PREFACE
Thank you for choosing the pH 510 pH/mV/Temperature or Ion 510 Ion/pH/mV/Temperature bench meter. The instruction manual functions in two ways: firstly as a step-by-step guide to help you operate the 510 series meter. Secondly, it serves as a handy reference guide. The instruction manual is written to cover as many anticipated applications of the pH/Ion 510 Bench meter series as possible. If there are doubts in the use of the instrument, do not hesitate to contact the nearest Authorized Distributor. Eutech Instruments/ Oakton Instruments cannot accept any responsibility for damage or malfunction to the meter caused by improper use of the instrument. The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on the part of Eutech Instruments Pte Ltd/ Oakton Instruments.
Note: Eutech Instruments Pte Ltd/ Oakton Instruments reserves the right to make improvements in design, construction, and appearance of our products without notice. Copyright © 2002 All rights reserved. Eutech Instruments Pte Ltd Oakton Instruments Rev 6 08/04
TABLE OF CONTENTS
1
INTRODUCTION ........................................................................................................................... 1
1.1 1.2 1.3 1.4 Introducing the 510 Bench Series ...............................................................................................................1 Custom LCD ................................................................................................................................................1 Keypad.........................................................................................................................................................2 Rear Instrument Panel.................................................................................................................................3
2
STARTING UP .............................................................................................................................. 4
2.1 2.2 2.3 2.4 2.5 2.6 Connecting the Sensor Electrode................................................................................................................4 Connecting the Temperature Probe ............................................................................................................4 Connecting the A/C Adapter........................................................................................................................4 Connecting the Chart Recorder...................................................................................................................4 Connecting the Electrode Holder (Optional) ...............................................................................................5 Steps Prior to Calibration.............................................................................................................................6
3
CALIBRATION .............................................................................................................................. 6
3.1 3.2 3.3 3.4 pH Calibration ..............................................................................................................................................6 Ion Concentration Calibration (For Ion 510 only) ........................................................................................9 mV Calibration ...........................................................................................................................................11 Temperature Calibration............................................................................................................................12
4
MEASUREMENT..........................................................................................................................13
4.1 4.2 4.3 4.4 4.5 Automatic Temperature Compensation (ATC)..........................................................................................13 Manual Temperature Compensation (MTC) .............................................................................................13 Taking Measurements ...............................................................................................................................14 HOLD Function ..........................................................................................................................................15 Storing and Recalling Data from Memory .................................................................................................15
5
SETUP FUNCTIONS....................................................................................................................17
5.1 5.2 5.3 5.4 P1.0 P2.0 P3.0 P4.0 Clear Memory...................................................................................................................................17 Viewing Electrode Diagnosis ..........................................................................................................18 Selecting pH Buffer Standard ..........................................................................................................19 Meter Reset......................................................................................................................................19
6
pH Electrode Maintenance .........................................................................................................20
6.1 6.2 6.3 6.4 Storage ......................................................................................................................................................20 After Use ....................................................................................................................................................20 Electrode Cleaning ....................................................................................................................................21 pH Electrode Rejuvenation........................................................................................................................22
7
TROUBLESHOOTING GUIDE .....................................................................................................23
7.1 7.2 Error Messages .........................................................................................................................................23 Troubleshooting .........................................................................................................................................24
8
ADDITIONAL INFORMATION .....................................................................................................25
8.1 8.2 8.3 8.4 8.5 pH and Temperature .................................................................................................................................25 Use of Standard pH Buffers ......................................................................................................................25 Standard pH Buffers ..................................................................................................................................25 List of Accessories.....................................................................................................................................26 Specifications.............................................................................................................................................28
9
WARRANTY.................................................................................................................................29
10 RETURN OF ITEMS .....................................................................................................................29
Instruction Manual
pH 510/ Ion 510
1 INTRODUCTION
1.1 Introducing the 510 Bench Series
The pH 510 and Ion 510 bench meters are microprocessor-based which incorporates new ASIC (Application Specific Integrated Circuit). It is designed with convenience in mind, and offers many advanced user-friendly features. The meters are capable of storing and recalling up to 50 data sets in its non-volatile memory. In addition, as a space saver, an optional swivel electrode holder can be attached at the either side of bench meter for resting the electrodes and probes during operation. The pH 510 meter is capable of measuring pH/ORP with Temperature; while the Ion 510 meter measures Ion Concentration/pH/ORP with temperature. It is equipped with a large customized LCD (Liquid Crystal Display) with simultaneous display of the measured values for easy reading. It is most ideal for routine pH/Ion Concentration measurement in indoor applications.
1.2
Custom LCD
The 510 meter is characterized by a large custom dual LCD. The display has mode annunciators for pH, mV, ppm and temperature readings. The secondary (lower) display shows the temperature reading simultaneously with the primary (upper) display that shows the measured values of pH, ppm or mV. Special annunciators such as units of measure, error message, graphical icons and modes of operation are arranged around both primary and secondary display to give you a comprehensive yet uncluttered display.
SETUP Â The meter is in SETUP mode.
MEAS- The instrument is in the measurement mode.
CAL - The instrument is in the calibration mode.
MEM- Data is stored into memory
READY- The measured value has stabilized.
SETUP READY HOLD
MEAS
CAL
MEM
R.mV pH ppm
HOLD- The displayed value has been frozen.
Shows the respective mode of operation.
ERR - An error in operation has occured.
ERR
CF pH ATC
Buffer - Prompts user to select buffer during calibration, or indicates buffer error when flashing together with the ERR annunciator.
ELECTRODE- indicates electrode error when it appears with ERR.
pH Â Indicates buffer values of respective USA or NIST buffer option set.
ATC (Automatic Temperature Compensation) activated. If ATC is not displayed, temperature probe is not connected. In this case the Manual Temperature Compensation is activated.
1
Instruction Manual
pH 510/ Ion 510
1.3
Keypad
A large splash proof membrane keypad with tactile feedback makes meter easy to use. Names and symbols describe the function button controls. Key ON/OFF Measurement Mode Powers the meter ON/OFF. When the meter is switched on, it automatically starts in the last mode of operation. Toggles between the measurement and calibration modes of the meter. In SETUP mode, pressing CAL/MEAS key returns to the measurement mode. Toggles between the different measurement modes available pH/Temperature/mV/Ion. (Ion mode only applies with the ION 510.) MI inputs measurements into memory and scrolls through memory values. Calibration / Setup Mode ~
CAL/MEAS
~
MODE
Switches from pH to Temperature in pH calibration mode scrolls values in mV calibration mode and selects temperature values and ion options in calibration mode. Scrolls to next program in the SETUP mode. scrolls values in mV calibration mode and selects temperature values and ion options in calibration Scrolls to next program in the SETUP mode. ~ Confirms and enters the value selected for calibration.
MI/
MR/
MR recalls measurements from memory and scrolls through memory values.
HOLD ENTER
Freezes a measurement on the display. Press again to get current readings. Enter functions in the memory mode.
You can also configure functions like Meter Reset, Memory Clear and Selection of pH Buffer Standard via the keypad. Please refer to Section 5 for more information on the above functions.
2
Instruction Manual
pH 510/ Ion 510
1.4
Rear Instrument Panel
This 510 series meter provides a complete set of connectors for the various accessories commonly used. Listed in the table are the details of the connections that you can make. Rear View of Meter
Connection ATC INPUT REF REC GND DC
Function For phono jack connection from the temperature probe for Automatic Temperature compensation. The probe should be a 30K thermistor type. For connection to sensor electrodes with BNC type connectors. The meter accepts any pH, ORP or ISE with BNC connector. Always ensure that the connector is clean and dry. For connection to pin type reference electrode normally used with half cell (mono) type pH electrodes or ISEs. For connection to strip chart recorders. Use subminiature plug with positive tip. For connection to the ground earth jack (standard tip connectors). For connection for the AC/DC power adapter
3
Instruction Manual
pH 510/ Ion 510
2 STARTING UP
Connect the accessory connectors at the rear of the instrument panel. During operation, it is important that water does not get onto the BNC connector. Also avoid touching the connector with soiled or wet hands.
2.1
Connecting the Sensor Electrode
Slide the electrode connector of the electrode over the socket of ...