Detailed instructions for use are in the User's Guide.
Open this leaflet for calibration/testing/maintenance instructions.
Changing Batteries
1. Open battery INC compartment lid (with attached lanyard loop). 2. Remove old batteries; replace DEC with fresh ones. Note Diagram 1 polarity as shown in battery compartment. See Diagram 1. 3. Recalibrate after battery change.
Warranty
Eutech Instruments warrants its meter free from manufacturing defects for 2 years and electrode module for 6 months. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period, please return the tester  freight prepaid  and correction will be made without charge. Out of warranty products will be repaired on a charge basis.
Return Of Items
Authorization must be obtained from Eutech's distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned. Note: We reserve the right to make improvements in design, construction and appearance of products without notice. Prices are subject to change without notice.
+ +
· Waterproof & Floats · Replaceable Electrode · Auto-off · Automatic Temperature Compensation (ATC)
Waterproof ECScan High Instructions
Before You Begin
Remove electrode cap. Soak electrode for a few minutes in alcohol to remove oils.
Conductivity Testing
1. Remove electrode cap. Switch unit on (ON/OFF key). 2. Dip electrode into test solution. Make sure sensor is fully covered. 3. Wait for reading to stabilize (Automatic Temperature Compensation cor rects for temperature changes). Note reading. 4. Press ON/OFF to turn off tester. Replace electrode cap. Note: Tester automatically shuts off after 8.5 minutes of non-use.
Calibration
Tester is factory calibrated. However, to ensure accuracy, calibrate conductivity on a regular basis. Select a calibration standard appropriate for the ECScan High tester (between 2.00 mS and 19.90 mS). It is best to select a standard close to the test solution value. 1. Open battery compartment lid (end with lanyard loop). 2. Orientate the battery compartment as shown in Diagram 1. 3. Rinse electrode in deionized water, then rinse it in calibration standard, then dip it into a container of calibration standard. 4. Switch unit on (ON/OFF key). Wait several minutes for display to stabilize. 5. Press the INC or DEC key to adjust reading to match the calibration standard value. 6. After 3 seconds without a key press, the display flashes 3 times, then shows `ENT'. The tester accepts calibration value; returns to measurement mode. 7. Replace battery cap.
HOLD Function
Press HOLD key to freeze display. Press HOLD again to release.
Tester Maintenance
· To improve performance, clean the electrode by rinsing them in alcohol for 10 to 15 minutes. · Replace all batteries if low battery indicator appears, or if readings are faint or unstable. · If you experience drift, periodically let electrode fully dry. When you need a new electrode, see "Electrode Replacement" on insert in back of box.
Self-Diagnostic Messages
bAt Weak batteriesÂreplace with fresh ones specified by manufacturer. Err Wrong keypress. Or, Ur Over range/Under range signal, or electrode is not in contact with solution, or electrode is failing. ---- Calibration mode activated but not performed.
68X247755 11/00 Rev 0
Open this leaflet for calibration/testing/maintenance instructions.
Changing Batteries
1. Open battery INC compartment lid (with attached lanyard loop). 2. Remove old batteries; replace with fresh ones. DEC Note polarity Diagram 1 as shown in battery compartment. See Diagram 1. 3. Recalibrate after battery change.
Warranty
Eutech Instruments warrants its meter free from manufacturing defects for 2 years and electrode module for 6 months. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period, please return the tester  freight prepaid  and correction will be made without charge. Out of warranty products will be repaired on a charge basis.
Return Of Items
Authorization must be obtained from Eutech's distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned. Note: We reserve the right to make improvements in design, construction and appearance of products without notice. Prices are subject to change without notice.
+
· Waterproof & Floats · Replaceable Electrode · Auto-off · Automatic Temperature Compensation (ATC)
Waterproof ECScan Low Instructions
Before You Begin
Remove electrode cap. Soak electrode for a few minutes in alcohol to remove oils.
Conductivity Testing
1. Remove electrode cap. Switch unit on (ON/OFF key). 2. Dip electrode into test solution. Make sure sensor is fully covered. 3. Wait for reading to stabilize (Automatic Temperature Compensation corrects for temperature changes). Note reading. 4. Press ON/OFF to turn off tester. Replace electrode cap. Note: Tester automatically shuts off after 8.5 minutes of non-use.
Calibration
Tester is factory calibrated. However, to ensure accuracy, calibrate conductivity on a regular basis. Select a calibration standard appropriate for the ECScan Low tester (between 200 µS and 1990 µS). It is best to select a standard close to the test solution value. 1. Open battery compartment lid (end with lanyard loop). 2. Orientate the battery compartment as shown in Diagram 1. 3. Rinse electrode in deionized water, then rinse it in calibration standard, then dip it into a container of calibration standard. 4. Switch unit on (ON/OFF key). Wait several minutes for display to stabilize. 5. Press the INC or DEC key to adjust reading to match the calibration standard value. 6. After 3 seconds without a key press, the display flashes 3 times, then shows `ENT'. The tester accepts calibration value; returns to measurement mode. 7. Replace battery cap.
HOLD Function
Press HOLD key to freeze display. Press HOLD again to release.
Tester Maintenance
· To improve performance, clean the electrode by rinsing them in alcohol for 10 to 15 minutes. · Replace all batteries if low battery indicator appears, or if readings are faint or unstable. · If you experience drift, periodically let electrode fully dry. When you need a new electrode, see "Electrode Replacement" on insert in back of box.
Self-Diagnostic Messages
bAt Weak batteriesÂreplace with fresh ones specified by manufacturer. Err Wrong keypress. Or, Ur Over range/Under range signal, or electrode is not in contact with solution, or electrode is failing. ---- Calibration mode activated but not performed.
68X247754 11/00 Rev 0
Twist-off Battery Compartment
REMOVE THE OLD ELECTRODE A
(Picture A)
Lanyard connection
Replaceable Sensor
REPLACE THE ELECTRODE B
(Picture B)
· Push-button calibration · Fast, stable, repeatable readings · Available through authorized distributors
SPECIFICATIONS
Tester Range TDScan TDScan Low High 0 to 1990 ppm 10 ppm ECScan ECScan Low High 0 to 0 to 0 to 10.00 1990 19.90 ppt µS/cm mS/cm 0.10 10 0.10 ppt µS/cm mS/cm ±1% Full Scale 0.4 to 1.0 selectable 200 to 2 .00 to 200 to 2.00 to 1990 10.00 1990 19.90 ppm ppt µS/cm mS/cm 1 point (calibration range is ±30% of factory default parameter) 0 to 50 °C 2% per °C 0 to 50 °C Four X 1.5V (Type: A76, LR 44 or V136A); > 140 hrs continuous use Tester: 16.5 cm x 3.8 cm; 90 g Boxed: 22 cm x6 cm x 5 cm; 170 g Manufactured By:
Electrode Replacement
You can replace the electrode module at the fraction of the cost of a new tester. When the tester fails to calibrate or gives fluctuating readings in calibration standards, you need to change the electrode.
1. With dry hands, grip the ribbed tester collar with electrode facing you. Twist the collar counter clockwise (see picture A). Save the ribbed tester collar and O-ring for later use. 2. Pull the old electrode module away from the tester. 3. Align the four tabs on the new module so that they match the four slots on the tester (see picture B). 4. Gently push the module onto the slots to sit it in position. Push the smaller O-ring fully onto the new electrode module. Push the collar over the module and thread it into place by firmly twisting clockwise.
Resolution Accuracy TDS factor Calibration Standard Range Calibration ATC Temp Coefficient Operating Temp Power Battery Dimension; Weight
Blk 55, Ayer Rajah Crescent, #04-14/24, Singapore 139949 Tel: (65)778 6876 Fax: (65)773 0836 E-Mail: marketing@eutechinst.com Web-Site: http://www.eutechinst.com
Eutech Instruments Pte Ltd
Applications
Water quality testing · water and Wastewater treatment · hydroponics · pools · boilers · factories · ecology studies · and more!
Made in Singapore