Detailed instructions for use are in the User's Guide.
Open this leaflet for calibration/testing/maintenance instructions.
Changing Batteries
1. Open battery INC compartment lid (with attached lanyard loop). 2. Remove old batteries; replace DEC with fresh ones. Note Diagram 1 polarity as shown in battery compartment. See Diagram 1. 3. Recalibrate after battery change.
Warranty
Eutech Instruments warrants its meter free from manufacturing defects for 2 years and electrode module for 6 months. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period, please return the tester  freight prepaid  and correction will be made without charge. Out of warranty products will be repaired on a charge basis.
Return Of Items
Authorization must be obtained from Eutech's distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned. Note: We reserve the right to make improvements in design, construction and appearance of products without notice. Prices are subject to change without notice.
+ +
· Waterproof & Floats · Replaceable Electrode · Auto-off · Automatic Temperature Compensation (ATC)
Waterproof ECScan High Instructions
Before You Begin
Remove electrode cap. Soak electrode for a few minutes in alcohol to remove oils.
Conductivity Testing
1. Remove electrode cap. Switch unit on (ON/OFF key). 2. Dip electrode into test solution. Make sure sensor is fully covered. 3. Wait for reading to stabilize (Automatic Temperature Compensation cor rects for temperature changes). Note reading. 4. Press ON/OFF to turn off tester. Replace electrode cap. Note: Tester automatically shuts off after 8.5 minutes of non-use.
Calibration
Tester is factory calibrated. However, to ensure accuracy, calibrate conductivity on a regular basis. Select a calibration standard appropriate for the ECScan High tester (between 2.00 mS and 19.90 mS). It is best to select a standard close to the test solution value. 1. Open battery compartment lid (end with lanyard loop). 2. Orientate the battery compartment as shown in Diagram 1. 3. Rinse electrode in deionized water, then rinse it in calibration standard, then dip it into a container of calibration standard. 4. Switch unit on (ON/OFF key). Wait several minutes for display to stabilize. 5. Press the INC or DEC key to adjust reading to match the calibration standard value. 6. After 3 seconds without a key press, the display flashes 3 times, then shows `ENT'. The tester accepts calibration value; returns to measurement mode. 7. Replace battery cap.
HOLD Function
Press HOLD key to freeze display. Press HOLD again to release.
Tester Maintenance
· To improve performance, clean the electrode by rinsing them in alcohol for 10 to 15 minutes. · Replace all batteries if low battery indicator appears, or if readings are faint or unstable. · If you experience drift, periodically let electrode fully dry. When you need a new electrode, see "Electrode Replacement" on insert in back of box.
Self-Diagnostic Messages
bAt Weak batteriesÂreplace with fresh ones specified by manufacturer. Err Wrong keypress. Or, Ur Over range/Under range signal, or electrode is not in contact with solution, or electrode is failing. ---- Calibration mode activated but not performed.
68X247755 11/00 Rev 0
Open this leaflet for calibration/testing/maintenance instructions.
Changing Batteries
1. Open battery INC compartment lid (with attached lanyard loop). 2. Remove old batteries; replace with fresh ones. DEC Note polarity Diagram 1 as shown in battery compartment. See Diagram 1. 3. Recalibrate after battery change.
Warranty
Eutech Instruments warrants its meter free from manufacturing defects for 2 years and electrode module for 6 months. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period, please return the tester  freight prepaid  and correction will be made without charge. Out of warranty products will be repaired on a charge basis.
Return Of Items
Authorization must be obtained from Eutech's distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned. Note: We reserve the right to make improvements in design, construction and appearance of products without notice. Prices are subject to change without notice.
+
· Waterproof & Floats · Replaceable Electrode · Auto-off · Automatic Temperature Compensation (ATC)
Waterproof ECScan Low Instructions
Before You Begin
Remove electrode cap. Soak electrode for a few minutes in alcohol to remove oils.
Conductivity Testing
1. Remove electrode cap. Switch unit on (ON/OFF key). 2. Dip electrode into test solution. Make sure sensor is fully covered. 3. Wait for reading to stabilize (Automatic Temperature Compensation corrects for temperature changes). Note reading. 4. Press ON/OFF to turn off tester. Replace electrode cap. Note: Tester automatically shuts off after 8.5 minutes of non-use.
Calibration
Tester is factory calibrated. However, to ensure accuracy, calibrate conductivity on a regular basis. Select a calibration standard appropriate for the ECScan Low tester (between 200 µS and 1990 µS). It is best to select a standard close to the test solution value. 1. Open battery compartment lid (end with lanyard loop). 2. Orientate the battery compartment as shown in Diagram 1. 3. Rinse electrode in deionized water, then rinse it in calibration standard, then dip it into a container of calibration standard. 4. Switch unit on (ON/OFF key). Wait several minutes for display to stabilize. 5. Press the INC or DEC key to adjust reading to match the calibration standard value. 6. After 3 seconds without a key press, the display flashes 3 times, then shows `ENT'. The tester accepts calibration value; returns to measurement mode. 7. Replace battery cap.
HOLD Function
Press HOLD key to freeze display. Press HOLD again to release.
Tester Maintenance
· To improve performance, clean the electrode by rinsing them in alcohol for 10 to 15 minutes. · Replace all batteries if low battery indicator appears, or if readings are faint or unstable. · If you experience drift, periodically let electrode fully dry. When you need a new electrode, see "Electrode Replacement" on insert in back of box.
Self-Diagnostic Messages
bAt Weak batteriesÂreplace with fresh ones specified by manufacturer. Err Wrong keypress. Or, Ur Over range/Under range signal, or electrode is not in contact with solution, or electrode is failing. ---- Calibration mode activated but not performed.
68X247754 11/00 Rev 0
Open this leaflet for calibration/testing/maintenance instructions.
Changing Batteries
1. Open battery INC compartment lid (with attached lanyard loop). 2. Remove old batteries; replace with fresh ones. Note polarity DEC as shown Diagram 1 in battery compartment. See Diagram 1. 3. Recalibrate after battery change.
Tester Maintenance
· To improve performance, clean the electrode by rinsing them in alcohol for 10 to 15 minutes. · Replace all batteries if low battery indicator appears, or if readings are faint or unstable. · If you experience drift, periodically let electrode fully dry. When you need a new electrode, see "Electrode Replacement" on insert in back of box.
Warranty
Eutech Instruments warrants its meter free from manufacturing defects for 2 years and electrode module for 6 months. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period, please return the tester  freight prepaid  and correction will be made without charge. Out of warranty products will be repaired on a charge basis.
+ +
Return Of Items
Authorization must be obtained from Eutech's distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned. Note: We reserve the right to make improvements in design, construction and appearance of products without notice. Prices are subject to change without notice.
· Waterproof & Floats · Replaceable Electrode · Auto-off · Adjustable TDS Factor · Automatic Temperature Compensation (ATC)
WP TDScan High Instructions
Before You Begin
Remove electrode cap. Soak electrode for a few minutes in alcohol to remove oils.
TDS Testing
1. Remove electrode cap. Switch unit on (ON/OFF key). 2. Dip electrode into test solution. Make sure sensor is fully covered. 3. Wait for reading to stabilize (Automatic Temperature Compensation corrects for temperature changes). Note reading. 4. Press ON/OFF to turn off tester. Replace electrode cap. Note: Tester automatically shuts off after 8.5 minutes of non-use.
Calibration
Tester is factory calibrated. However, to ensure accuracy, calibrate TDS on a regular basis. Select a calibration standard appropriate for the TDScan High tester (between 2.00 ppt and 10.00 ppt). It is best to select a standard close to the test solution value. 1. Open battery compartment lid (end with lanyard loop). 2. Orientate the battery compartment as shown in Diagram 1. 3. Rinse electrode in deionized water, then rinse it in calibration standard, then dip it into a container of calibration standard. 4. Switch unit on (ON/OFF key). Wait several minutes for display to stabilize. 5. Press the INC or DEC keys to adjust reading to match the calibration standard value. 6. After 3 seconds without a key press, the display flashes 3 times, then shows `ENT'. The tester accepts calibration value; returns to measurement mode. 7. Replace battery cap.
HOLD Function
Press HOLD key to freeze display. Press HOLD again to release.
Setting TDS Factor
This tester lets you select a TDS factor of 0.4 to 1.0. 1. Open battery compartment. With meter on, press the HOLD key, then press the INC key See Diagram 1. 2. Press the INC or DEC key to adjust the TDS factor. 3. After 3 seconds without a key press, the display flashes 3 times, then shows `ENT'. Tester accepts TDS factor and returns to measurement mode. 4. Replace battery cap.
Self-Diagnostic Messages
bAt Weak batteriesÂreplace with fresh ones specified by manufacturer. Err Wrong keypress. Or, Ur Over range/Under range signal, or electrode is not in contact with solution, or electrode is failing. ---- Calibration mode activated but not performed.
68X247753 11/00 Rev 0
Open this leaflet for calibration/testing/maintenance instructions.
Changing Batteries
1. Open battery INC compartment lid(with attached lanyard loop). 2. Remove old batteries; replace with fresh ones. Note polarity DEC as shown Diagram 1 in battery compartment. See Diagram 1. 3. Recalibrate after battery change.
Tester Maintenance
· To improve performance, clean the electrodes by rinsing them in alcohol for 10 to 15 minutes. · Replace all batteries if low battery indicator appears, or if readings are faint or unstable. · If you experience drift, periodically let electrode fully dry. When you need a new electrode, see "Electrode Replacement" on insert in back of box.
Warranty
Eutech Instruments warrants its meter free from manufacturing defects for 2 years and electrode module for 6 months. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period, please return the tester  freight prepaid  and correction will be made without charge. Out of warranty products will be repaired on a charge basis.
+
Return Of Items
Authorization must be obtained from Eutech's distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned. Note: We reserve the right to make improvements in design, construction and appearance of products without notice. Prices are subject to change without notice.
· Waterproof & Floats · Replaceable Electrode · Auto-off · Adjustable TDS Factor · Automatic Temperature Compensation (ATC)
WP TDScan Low Instructions
Before You Begin
Remove electrode cap. Soak electrode for a few minutes in alcohol to remove oils.
TDS Testing
1. Remove electrode cap. Switch unit on (ON/OFF key). 2. Dip electrode into test solution. Make sure sensor is fully covered. 3. Wait for reading to stabilize (Automatic Temperature Compensation corrects for temperature changes). Note reading. 4. Press ON/OFF to turn off tester. Replace electrode cap. Note: Tester automatically shuts off after 8.5 minutes of non-use.
Calibration
Tester is factory calibrated. However, to ensure accuracy, calibrate TDS on a regular basis. Select a calibration standard appropriate for the TDScan Low tester (between 200 ppm and 1990 ppm). It is best to select a standard close to the test solution value. 1. Open battery compartment lid (end with lanyard loop). 2. Orientate the battery compartment as shown in Diagram 1. 3. Rinse electrode ...