Instruction Manual
HI 8427 - HI 931001
pH/mV Simulators
pH/m HI 8427 V CA LIBR ATO
R
pH2
LOW BATT ERY LEAK AGE pH4 pH7
pH0 OFF
pH10
pH12 pH14 -350 mV 350 mV
1900 mV
0 50 40 °C
10 20 30
1900 mV
HIGH IMPE DA
NCE
These Instruments are in Compliance with the CE Directives
1
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a more precise idea of its versatility in a wide range of applications. These instruments are in compliance with the directives EN 50081-1 and EN 50082-1.
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If noticeable damage is found, immediately notify your Dealer. Each pH/mV simulator is supplied complete with: · HI 7858/1 BNC/BNC Coaxial Cable; Note: Save all packing material until the instrument has been observed to function correctly because all defective items must be returned to the Dealer in their original packing.
TABLE OF CONTENTS Preliminary Examination ........................... 3 General Description .................................. 3 Functional Description HI 8427 ................. 4 Functional Description HI 931001 ............. 5 pH Calibration ........................................... 6 mV (ORP) Calibration ............................... 8 High Impedance Calibration .................... 10 pH Values at Various Temperature .......... 11 Battery Replacement ............................... 12 Accessories ............................................. 13 Warranty .................................................. 14 CE Declaration of Conformity ................. 15 GENERAL DESCRIPTION HI 8427 and HI 931001 are portable pH/mV simulators designed specifically for calibrating pH and mV (ORP) meters. Two calibration ranges are provided: from 0 to 14 pH and from -1900 to 1900 mV with HI 8427; from 0 to 14 pH and from -1000 to 1000 mV with HI 931001. HI 931001 provides with pH/mV values at 25°C. HI 8427 allows the manual temperature compensation and is equipped with a HIGH IMPEDANCE key and LEAKAGE LED to verify if the connection cable is in perfect condition. A low battery indicator will be lit when the simulator battery runs low.
3
ISO 9000 Certified Company since 1992
2
FUNCTIONAL DESCRIPTION HI 8427
9
FUNCTIONAL DESCRIPTION HI 931001
HI 8427 pH/mV CALIBRATOR
7 8
pH2
LOW BATTERY LEAKAGE pH4 pH7 pH10
3 1 2
pH0 pH12
3
OFF
pH14
1900 mV
-350 mV
350 mV
1900 mV
4 6
10 0
5
50 40 °C
20 30 HIGH IMPEDANCE
1. 2. 3. 4. 5. 6. 7. 8. 9.
Rotary Knob Switch Off Selection pH Outputs Selection mV Outputs Selection Temperature Compensation Knob High Impedance Key Low Battery LED Leakage LED BNC Socket
pH mV pH mV 0, 2, 4, 7, 10, 12, 14 -1900, -350, 350, 1900 ±0.1 ±5 Manual from 0 to 50°C (32 to 122°F) 109 9V 100 hours of continuous use 0 to 50°C (32 to 122°F); 95% RH 180x83x40 mm (7.1 x 3.3 x 1.6") 255 g (9 oz.)
1. 2. 3. 4. 5. 6.
LCD Display ON/OFF Key Range Selection Key pH/mV Coarse Setting Knob pH/mV Fine Setting Knob BNC Socket
pH mV pH mV 0.00 to 14.00 -1000 to 1000 0.01 1 ±0.01 ±1 ±0.02 ±2 All output values simulated at 25°C (77°F) 9V 500 hours of continuous use 0 to 50°C (32 to 122°F); 95% RH 180 x 83 x 40 mm (7.1 x 3.3 x 1.6") 320 g (11.3 oz.)
Range Resolution
Range Accuracy (@20°C/68°F) Temperature Compensation
Accuracy pH (@20°C/68°F) mV EMC Typical pH Deviation mV Temperature Compensation Battery Type Life Environment Dimensions Weight
High Impedance Test Battery Type Life Environment Dimensions Weight
4
5
pH CALIBRATION
HI 8427: · Connect the simulator to the pH meter using the HI 7858/1 cable provided.
pH
working temperature. If not, adjust the pH meter to read the correct value. Note: The °C knob is not effective during pH 7 calibration, since pH 7 is used as reference.
HI 8427 pH/mV CALIBRATOR
·
·
·
Switch the pH meter on and select pH 7 on the simulator. Make sure that the LOW BATTERY LED is not lit. If it is lit, change the battery of the simulator before performing the calibration. Verify that the pH meter displays the pH 7 message. If not, adjust the pH meter until the correct reading pH 7 is displayed (refer to the pH meter instruction manual). Select any other pH value on the simulator and set the °C knob to the desired working temperature (e.g. pH 4 at 30°C).
pH7
HI 931001: · Connect the simulator to the pH meter using the HI 7858/1 cable provided. · Manually set the temperature compensation of your meter to 25°C.
· Switch the simulator on and select the pH mode by pressing the RANGE key.
pH
pH
HI 8424
microcomputer pHmeter
ON OFF
HI 931001
pH/mV Calibrator
SETTING
ON OFF
RANGE
pH
HI 8427 pH/mV CALIBRATOR
·
0
10 20 30
Make sure that the two additional decimal points are not displayed. If they are displayed, change the battery of the simulator before performing the calibration. Turn the Coarse and Fine knob to set any desired pH value and make sure that the pH meter you are checking reflects the correct result (e.g. 7.01 pH).
·
50 40 °C
·
·
If your pH meter has a temperature probe connected to it, set the °C knob to the measured temperature displayed by your pH meter. If your meter is manually compensated for temperature, adjust the temperature settings for both the pH meter and the simulator to the same temperature. Verify that the pH meter displays the correct pH value compensated at the
6
·
If not, adjust the pH meter until the correct reading is displayed (refer to the pH meter instruction manual).
pH
pH
HI 8424
microcomputer pHmeter
ON OFF
HI 931001
pH/mV Calibrator
SETTING
7
mV (ORP) CALIBRATION
·
HI 8427: · Connect the simulator to the mV (ORP) meter using the HI 7858/1 cable provided.
·
·
pH
HI 8427 pH/mV CALIBRATOR
Switch the mV meter on and select the desired mV value on the simulator. Make sure that the LOW BATTERY LED is not lit. If it is lit, change the bat-350 tery of the simulator bemV fore performing the calibration. · If the meter is not displaying the correct mV value, adjust the meter until it displays the correct mV (refer to the mV meter instructions manual). Note: The °C knob is not effective during mV calibration.
Make sure that the the two additional decimal points are not displayed. If they are displayed, change the battery of the simulator before performing the calibration. Turn the Coarse or Fine knob to any desired value and make sure that the mV meter you are checking reflects the correct result.
·
If not, adjust the mV meter until the correct reading is displayed (refer to the mV meter instruction manual).
mV
mV
HI 8424
microcomputer pHmeter
ON OFF
HI 931001
pH/mV Calibrator
SETTING
Note: Temperature compensation does not apply here.
HI 931001: · Connect the simulator to the mV (ORP) meter using the HI 7858/1 cable provided.
pH
pH
HI 8424
microcomputer pHmeter
ON OFF
HI 931001
pH/mV Calibrator
SETTING
ON
· Switch the simulator on and select the mV mode by pressing the RANGE key.
OFF
RANGE
8
9
CABLE HIGH IMPEDANCE TEST
pH VALUES AT VARIOUS TEMPERATURE
HI 8427 only:
Temperature has an effect on pH. The calibration buffer solutions are effected by temperature changes to a lesser degree than normal solutions (see table below). This pH simulation will not be affected by this chemical behavior. Note: With HI 8427 your pH meter will always show the pH buffer solution at the standard temperature of 25ºC.
·
Connect the simulator to the cable (e.g.HI 7855).
·
Press the HIGH IMPEDANCE key and observe the LEAKAGE LED while you are turning the simulator rotary knob to any pH/mV value.
HIGH IMPEDANCE
·
If the LED does not light, the cable and connector going to the meter are in perfect condition. If the LED is lit, it means there is a leakage problem with the cable and connector and you should check/replace it.
LEAKAGE
TEMP °C °F
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 32 41 50 59 68 77 86 95 104 113 122 131 140 149 158
4.01
4.01 4.00 4.00 4.00 4.00 4.01 4.02 4.03 4.04 4.05 4.06 4.07 4.09 4.11 4.12
pH VALUES 6.86 7.01 9.18
6.98 6.95 6.92 6.90 6.88 6.86 6.85 6.84 6.84 6.83 6.83 6.84 6.84 6.85 6.85 7.13 7.10 7.07 7.04 7.03 7.01 7.00 6.99 6.98 6.98 6.98 6.98 6.98 6.99 6.99 9.46 9.39 9.33 9.27 9.22 9.18 9.14 9.10 9.07 9.04 9.01 8.99 8.97 8.95 8.93
10.01
10.32 10.24 10.18 10.12 10.06 10.01 9.96 9.92 9.88 9.85 9.82 9.79 9.77 9.76 9.75
·
LEAKAGE
10
11
BATTERY REPLACEMENT HI 710001 The instruments use a 9 volt battery that lasts for approximately 100 hours or 500 hours of continuous use with HI 8427 and HI 931001 respectively. In the event of dead battery the LOW BATTERY LED of HI 8427 will be lit or two additional decimal points will be displayed on HI 931001.
LOW BATTERY LEAKAGE
ACCESSORIES Soft carrying case for HI 931001, dimensions 230 x 100 x 50 mm Extension cable with 3 mm diameter (0.12"), 1m (3.3') long Extension cable with 3 mm diameter (0.12"), 3m (9.9') long Extension cable with 3 mm diameter (0.12"), 5m (16.5') long
HI 7858/1 HI 7858/3 HI 7858/5
pH
HI 7858/10 Extension cable with 3 mm diameter (0.12"), 10m (33') long HI 7858/15 Extension cable with 3 mm diameter (0.12"), 15m (49.5') long
HI 7858 SERIES CABLE CONNECTORS BNC AND 3.0 mm (0.12") CABLE WITH BNC
Battery replacement must only take place in a non hazardous area using the battery types specified in this instruction manual (see page 11). To change battery, slide the bottom back cover, replace the old battery and replace the cover.
CONNECT TO THE BNC SOCKET OF THE SIMULATOR
CONNECT TO THE BNC SOCKET OF THE METER
HI 710031
Hard carrying case, dimensions 340 x 230 x 90 mm
BATTERY
12
13
WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrodes and the probes are warranted for a period of six months. This warranty is limited to repair or replacement free of charge. Damages due to accidents, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. In HI 8427, with an applied field of 3V/m, no leakage is signaled by the LEAKAGE led. To avoid damages or burns, do not perform any measurement in microwave ovens.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
14
15
www.hannainst.com