Instruction Manual
HI 9023C - HI 9023CN HI 9024C - HI 9025C HI 9110 - HI 9210 HI 9210N Portable Waterproof pH Meters
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument. It will provide you with the necessary information for a correct use of the instrument, as well as a precise idea of its versatility. If you require further technical information, do not hesitate to e-mail us at tech@hannainst.com. directives. These instruments are in compliance with the
TABLE OF CONTENTS
PRELIMINARY EXAMINATION ..................................................... 3 GENERAL DESCRIPTION ............................................................ 3 FUNCTIONAL DESCRIPTION HI 9023C & HI 9023CN .................. 5 SPECIFICATIONS HI 9023C & HI 9023CN ................................. 6 FUNCTIONAL DESCRIPTION HI 9024C & HI 9025C ..................... 7 SPECIFICATIONS HI 9024C & HI 9025C .................................... 8 FUNCTIONAL DESCRIPTION HI 9110.......................................... 9 FUNCTIONAL DESCRIPTION HI 9210 & HI 9210N .................... 10 SPECIFICATIONS HI 9110, HI 9210 & 9210N ......................... 11 OPERATIONAL GUIDE ............................................................. 12 pH CALIBRATION ................................................................... 16 pH VALUES AT VARIOUS TEMPERATURES ................................ 24 TROUBLESHOOTING GUIDE ..................................................... 25 ELECTRODE CONDITIONING AND MAINTENANCE......................... 26 TAKING REDOX MEASUREMENTS (HI 9023C & HI 9025C only) . 29 BATTERY REPLACEMENT & AUTO-OFF ...................................... 30 ACCESSORIES ....................................................................... 31 ELECTRODE APPLICATION REFERENCE GUIDE ............................. 37 WARRANTY .......................................................................... 38 CE DECLARATION OF CONFORMITY ......................................... 39
Manufacturers
since
1978
These Instruments are in Compliance with the CE Directives
2
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer immediately. Note: Save all packing materials until you are sure that the instrument functions correctly. Any damaged or defective item must be returned in its original packing materials together with the supplied accessories.
GENERAL DESCRIPTION
These waterproof Hanna instruments are heavy-duty pH meters designed to provide laboratory results and accuracy under harsh industrial conditions. HI 9023C, pH/mV/°C meter utilizes an advanced customized microprocessor. The meter has three memorized buffer values (4.01, 7.01 and 10.01) and automatic buffer recognition to avoid errors during calibration. There are no trimmers, making calibration an easy process, even for non-technical personnel. Temperature effects are automatically compensated for, or can be manually adjusted. With HI 9023C you can also measure ORP (Oxidation Reduction Potential) or Ion Specific. Millivolt measurements automatically switch from 0.1 to 1 mV resolution when the reading reaches 400mV. HI 9023C is supplied with HI 1230B combination double-junction gel pH electrode, HI 7669/2W temperature probe, pH 4.01 and 7.01 (20 mL each) buffer solutions, sample vessel, 1.5V AA size batteries (4 each) and a rugged carrying case. HI 9023CN, pH/mV/°C meter utilizes an advanced customized microprocessor. The meter has three memorized buffer values (4.01, 7.01 and 10.01) and automatic buffer recognition to avoid errors during calibration. Temperature effects are automatically compensated for. HI 9023CN is supplied with HI 1217D 4-in-1 amplified, gel pH/ C electrode, pH 4.01 and 7.01 (20 mL each) buffer solutions, sample vessel, 1.5V AA size batteries (4 each) and a rugged carrying case. HI 9024C and HI 9025C are pH/°C meters with a built-in microprocessor. A large dual-level LCD displays the pH and temperature simultaneously. The display has graphic symbols to make the calibration procedure easy to follow. The meters have 5 memorized buffer 3
values (4.01, 6.86, 7.01, 9.18 and 10.01), buffer recognition to avoid errors during calibration and automatic temperature compensation. You can use HI 9025C with ORP (Oxidation Reduction Potential) or Ion Specific Electrodes. Millivolt measurements automatically switch from 0.1 to 1 mV resolution when the reading reaches 400 mV. To insure trouble-free operation, the meter's circuitry comes with built-in protection against electromagnetic interference. HI 9024C and HI 9025C are supplied with HI 1230B combination double-junction, gel pH electrode, HI 7669/2W temperature probe, pH 4.01 and 7.01 (20 mL each) buffer solutions, sample vessel, 1.5V AA size batteries (4 each) and a rugged carrying case. HI 9110 pH meter has simple manual temperature compensation and calibration which make this meter easy to use in laboratory and field. The calibration trimmers are sealed behind a protective cover so that they cannot be accidentally moved or readjusted by unauthorized persons. HI 9110 is supplied with FC 100B combination double-junction, refillable pH electrode, calibration screwdriver, 1.5V AA size batteries (4 each) and a soft carrying case. HI 9210 pH meter has simple manual temperature compensation and calibration with only two buttons, making it simple to use for even non-technical personnel. The calibration trimmers are sealed behind a protective cover so that they cannot be accidentally moved or readjusted by unauthorized persons. HI 9210 is supplied with HI 1230B combination double-junction, gel pH electrode, calibration screwdriver and 1.5V AA size batteries (4 each). HI 9210N pH/°C meter measures both pH and temperature. It has only two buttons, making it simple to use for even non-technical personnel. The calibration trimmers are sealed behind a protective cover so that they cannot be accidentally moved or readjusted by unauthorized persons. HI 9210N is supplied with HI 1217D 4-in-1, gel, amplified pH/ C electrode, calibration screwdriver and 1.5V AA size batteries (4 each).
4
FUNCTIONAL DESCRIPTION HI 9023C & HI 9023CN
SPECIFICATIONS HI 9023C & HI 9023CN
HI 9023C Range HI 9023CN
1) 2) 3) 4) 5) 6) 7) 8) 9) 10)
Electrode connector (BNC for HI 9023C and DIN for HI 9023CN) Temperature probe socket (HI 9023C only) Liquid Crystal Display CAL key to enter or exit calibration mode CFM key to confirm calibration °C and C keys for manual temperature setting (HI 9023C only) ON/OFF key to turn the instrument on or off RANGE key to select pH, C or mV Low battery indicator (additional decimal point) Mode indicator 5
0.00 to 14.00 pH ±399.9 mV (ISE); ±1999 mV (ORP) 0.0 to 100.0 C 0.0 to 70.0 C Resolution 0.01 pH 0.1 mV (ISE); 1 mV (ORP) 0.1 C Accuracy ±0.01 pH (@20°C/68°F) ±0.2 mV (ISE); ±1 mV (ORP) ±0.4 C ±0.5 C Typical EMC ±0.02 pH Deviation ±0.2 mV; ±1 mV ±0.2 C ±0.5 C Calibration Automatic with 3 memorized standard buffers (4.01, 7.01, 10.01) Temperature Automatic or manual Automatic Compensation 0 to 100 C (32-212 F) 0 to 70 C (32-158 F) Electrode HI1230B plastic body, HI1217D plastic body, combination pH, BNC 4-in-1 gel pH/ C, DIN 1 m cable (included) 1 m cable (included) Temperature Probe HI 7669/2W (incl.) not necessary Input Impedance 1012 Ohm Battery Type & Life 4x1.5V, AA size (alkaline batteries) approx. 400 hours of continuous use auto-shut off after 10 minutes of non-use Environment 0 to 50 C (32 to 122 F); 100% RH Dimensions 196x80x60 mm (7.7x3.1x2.4") (meter) 340x230x80 mm (13.8x9x3.1") (kit) Weight 425 g (15 oz.) (meter) 1.3 Kg (3.0 lb.) (kit)
6
FUNCTIONAL DESCRIPTION HI 9024C & HI 9025C
SPECIFICATIONS HI 9024C & HI 9025C
HI 9024C Range HI 9025C
1) 2) 3) 4) 5) 6) 7) 8) 9) 10) 11) 12) 13)
BNC electrode connector Temperature probe socket Liquid Cristal Display RANGE key to select pH or mV (HI 9025C only) ON/OFF key to turn the meter on or off CAL key to enter or exit calibration mode CFM key to confirm calibration °C and C keys for manual temperature setting, or selecting pH buffer value MEM key to store pH value in memory MR key to recall the stored value from memory User-friendly graphic symbols Primary display Secondary display 7
0.00 to 14.00 pH; 0.0 to 100.0 C -- 0.0/±399.9 mV (ISE) -- ±400/±1999 mV (ORP) Resolution 0.01 pH; 0.1 C -- 0.1 mV (ISE) -- 1 mV (ORP) Accuracy ±0.01 pH; ±0.5 C (@20°C/68°F) -- ±0.2 mV (ISE) Typical EMC ±0.01 pH; ±0.1 C Deviation -- ±0.3 mV (ISE) Calibration Automatic with 5 memorized standard buffers (4.01, 6.86, 7.01, 9.18, 10.01) Offset Calibration ±1 pH Slope Calibration From 70 to 108% Temperature Automatic or manual from 0 to 100 C Compensation (32 to 212 F) Electrode HI 1230B pH electrode (included) Temperature Probe HI 7669/2W (included) Input Impedance 1012 Ohm Battery Type & Life 4x1.5V, AA size (alkaline batteries) approx. 400 hours of continuous use auto-shut off after 10 minutes of non-use Environment 0 to 50 C (32 to 122 F); 100% RH Dimensions meter: 196x80x60 mm (7.7x3.1x2.4") kit: 340x230x80 mm (13.8x9x3.1") Weight meter: 425 g (15 oz.) kit: 1.3 Kg (3.0 lb.)
8
FUNCTIONAL DESCRIPTION HI 9110
1 2 3
1
FUNCTIONAL DESCRIPTION HI 9210 & HI 9210N
3 4 5
4 5
2
7 6
pH
7 6
pH
Foodcare
pH METER
HI 9210N ATC pH METER
8
RANGE 0.00 TO 14.00 pH ACCURACY ±0.02pH
ON OFF pH °C SET
9
RANGE pH: 0.00 - 14.00 °C: 0.0 -70.0
8 9
ON/OFF
HI 9110
RANGE
1) 2) 3) 4) 5) 6) 7) 8) 9)
Battery cover BNC electrode connector Offset calibration trimmer Slope calibration trimmer Temperature setting trimmer Mode indicator Liquid Crystal Display ON/OFF key to turn the meter on or off pH/°C SET key to select pH measurement or for manual temperature setting 9
1) 2) 3) 4) 5) 6) 7) 8) 9)
Battery cover Electrode connector (BNC for HI 9210 and DIN for HI 9210N) Offset calibration trimmer Slope calibration trimmer Temperature setting trimmer (HI 9210) or Temperature Calibration (HI 9210N) Mode indicator Liquid Crystal Display ON/OFF key to turn the meter on or off RANGE key to visualize the temperature (HI 9210N) or for manual temperature setting (HI 9210) 10
SPECIFICATIONS HI 9110, HI 9210 & HI 9210N
HI 9110 Range HI 9210 HI 9210N
OPERATIONAL GUIDE
INITIAL PREPARATION
Each meter is supplied complete with batteries. Remove the back cover, unwrap the batteries and install them while paying attention to polarity.
pH 0.00 to 14.00 pH °C -- 0.0 to 70.0 C Resolution pH 0.01 pH °C -- 0.1 C Accuracy pH ±0.02 pH (@20°C/68°F) °C -- ±0.5 C Typical EMC pH ±0.05 pH ±0.02 pH Deviation °C -- ±0.5 C Calibration Manual 2 points through trimmers Offset Calibration ±1 pH Slope Calibration From 85 to 105% of nominal value Temperature Manual 0 to 100 C Auto 0 to 70 C Compensation (32 to 212 F) (32 to 158 F) Electrode FC 100B HI 1230B HI 1217D pH (incl.) pH (incl.) pH/ C (incl.) Input Impedance 1012 Ohm Battery Type 4x1.5 volt, AA size (alkaline batteries); Life 2000 hours of continuous use Environment 0 to 50 C (32 to 122 F); 100% RH Dimensions 196 x 80 x 60 mm (7.7 x 3.1 x 2.4") Weight 320 g (11.3 oz.)
HI 9023C, HI 9024C, HI 9025C: To prepare the instrument for use connect the pH electrode and the temperature probe to the BNC and temperature sockets on the top of the instrument. The temperature probe can be used independently to take temperature measurements, or it can be used in conjunction with the pH electrode to utilize the meter's ATC capability. If the probe is disconnected, temperature can also be set manually with the UP and DOWN keys. To or switch the instrument on, press and hold the ON/OFF key for a fraction of a second. The meter has a built-in protection against electromagnetic interference and the delayed response of the keys assures that the commands are not mistaken for stray signals.
°C °C
ON/OFF
HI 9110 and HI 9210: Connect the pH electrode to the BNC connector on the top of the instrument and press the ON/OFF key. HI 9210N and HI 9023CN:
ON/OFF
Connect the pH/ C electrode to the DIN connector on the top of the instrument and press the ON/OFF key.
ON/OFF
11
12
pH MEASUREMENTS
To take a pH measurement remove the electrode protective cap and simply submerge the tip (4cm/1½") of the electrode and the temperature probe (wherever applicable) into the sample to be tested.
HI 9023C, HI 9023CN and HI 9210N To use the meter's Automatic Temperature Compensation feature, simply submerge the pH/ C electrode into the sample and wait for a couple of minutes. The pH electrode also houses the temperature sensor which will measure and adjust for the temperature effect automatically. HI 9110, HI 9210: To use the meter's Manual Temperature Compensation feature: Press the pH/ C key of HI 9110 or the RANGE key of HI 9210 to select the manual temperature mode.
pH °C SET
Turn the instru ...