Detailed instructions for use are in the User's Guide.
Instruction Manual
HI 9032
Microprocessor Bench Conductivity/TDS/NaCl Fertilizer/°C/°F Meter with RS 232
These Instruments are in Compliance with the CE Directives
http://www.hannainst.com
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. This instrument is in compliance with CSA, UL (EN 50081-1 and EN 50082-1) direcand tives. TABLE OF CONTENTS Preliminary Examination ............................. 3 General Description ..................................... 3 The Front Panel .......................................... 5 Liquid Crystal Displays................................ 7 The Rear Panel ........................................... 8 Specifications ........................................... 10 Operational Guide ...................................... 11 Conductivity Calibration ............................. 21 NaCl Calibration ........................................ 31 Fertilizer Calibration .................................. 35 Setting Temperature .................................. 40 Setting Temperature Coefficient ................. 42 Setting TDS Factor ................................... 45 Programs .................................................. 47 Interface with PC ....................................... 54 Probe Maintenance ................................... 59 Accessories .............................................. 60 Warranty ................................................... 62 CE Declaration of Conformity .................... 63
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer. Each HI 9032 bench meter is supplied complete with: · HI 7669/2W Temperature Probe · HI 7686 Conductivity Probe · External Power Supply (HI710005 or HI710006) · HI 76405 Electrode Holder · Dust Cover Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original package together with the supplied accessories. GENERAL DESCRIPTION Hanna Instruments HI 9032 is a microprocessor-based, bench-top, conductivity meter designed for the laboratory stringent demands, with the ability to perform five types of measurements: · Conductivity (4 different ranges: 0.0 to 399.9 µS/cm, 399 to 3999 µS/cm, 3.99 to 39.99 mS/cm, 39.9 to 99.9 mS/cm) · Fertilizer Concentration (F1 or F2 in the range 0.00 to 39.9 g/l) · NaCl of sea water saturation (range 0.0 to 130.0%) · Temperature (range 0.0 to 50.0°C) · Total Dissolved Solids (3 different ranges: 0.0 to 199.9 ppm, 0 to 1999 ppm, 0.00 to 19.99 g/l).
3
ins rum n sCompany
ISO 9000 Certified since 1992
e
2
HI 9032 automatically recognizes buffers @ 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 80.0 mS/cm. Commands are easily accessible through the user-friendly keyboard: Through the COND key the meter is set in the conductivity range, TDS key selects Total Dissolved Solids measurements, NaCl key sets the range to NaCl measurement, and the FERT key selects 1st or 2nd fertilizer measurements. Temperature compensation can be either automatic or manual. It can be selected through the key ATC / MTC respectively. A temperature coefficient factor (for conductivity measurements) can be also selected through the %TC key (from 0.0 to 3.0%). The TDS factor (for TDS measurements) can be selected through the FTR key (from 0.01 to 0.99). The unit includes memory storages for all calibration data even after the unit is shut off. HI 9032 is equipped with two liquid crystal displays: · A large one in two sections (primary and secondary LCD) indicates conductivity, TDS, NaCl, Fertilizer and temperature coefficient %; · A smaller one indicates temperature. The meter also incorporates a standard R232C output to allow serial communication with a host computer; a series of programs can be selected for the setting of the R232C baud rate.
THE FRONT PANEL
CALIBRATION
1 Select Cond.
Fert.
NaCl
or
T.D.S.
ins rum n s
2 Depress
CAL
e
HI9032 microcomputer conductivity meter
3 Depress 0 .... 9 to enter calibration parameter 4 Depress ENTER repeat 3 & 4 when necessary
(refer to instruction manual for details)
3
key to confirm for
5 When 6 When
BUF
on L.C.D. is blinking, place electrode in reference buffer
displayed on L.C.D. & ~ on L.C.D. is blinking, equipment confirm that BUF buffer is within ±15% of theoritical set value, but reading is still not stable yet
BUF
1
LO BAT SAMPLE HOLD
mS µ ppm % gm
& " CON" displayed on L.C.D., depress CON accepting the calibration 8 Equipment return to OPERATION mode for measurement
7 When
CONDUCTIVITY TDS FERTILIZER NaCl
CON CAL
BUF
F1 F2
°C
%TC
Main Display Sub. Display Temp. Display
: The Result of measurement : Conditions of the measurement : Temperature reading (Keyboard entry is shown by the blinking of Decimal point)
ON
Cond. Fert. NaCl
OFF
4
CAL
2
TEMP .
HI LO
7 T.D.S.
8
9
4 FTR
5 °F
6 °C
CON
1 ATC MTC 0 2 %TC 3 PGM
ENTER
.
Clear
1. Large display for Conductivity, Fertilizer, NaCl and TDS This large LCD displays the conductivity, fertilizer, NaCl and TDS reading. The corresponding symbol will be displayed along with the measured value. A "----" signal will appear on this display to indicate an out of range condition. 2. Temperature display This LCD displays the temperature measured by the HI7669/2W probe or the value that was manually set through the keyboard. Temperature can be selected between the Celsius and the Fahrenheit scale by pressing the °C or the °F key respectively. If the probe is not connected when the meter is turned on, the "25°C" (or "77°F") symbol with a blinking decimal point will default.
5
4
If the temperature probe is disconnected during operation, the decimal point will start blinking. It will blink also if the temperature range is exceeded. This LCD also displays the program # during programming. 3. Calibration Procedure Briefly describes calibration procedure 4. Keyboard · Cond key, to select conductivity measurement mode · Fert. key, to select fertilizer measurement mode and to select Fertilizer #1 or #2 ("F1" or "F2") · NaCl key, to select NaCl measurement mode · T.D.S. key, to select total dissolved solids measurement mode · FTR key, to select the TDS factor for TDS measurements (see page 45) · °C key, to display temperature in the Celsius scale · °F key, to display temperature in the Fahrenheit scale · ATM/MTC key, to select automatic or manual temperature compensation
· ON key, to switch the instrument on · OFF key, to switch the instrument off · CAL key, to enter or exit the calibration mode · CON key, to confirm calibration data · ENTER Key, to confirm/enter the numeric data entry · 0 to 9 keys, numeric and decimal keys for numeric data input · Clear Key, to clear the numeric data entry LIQUID CRYSTAL DISPLAYS Large display for Conductivity, Fertilizer, NaCl and TDS
Primary Display
LO BAT SAMPLE HOLD
mS µ ppm % gm
CONDUCTIVITY TDS FERTILIZER NaCl
CON CAL
BUF
F1 F2
°C
%TC
Secondary Display
HI
· %T.C. key, to select the temperature coefficient for conductivity measurements or to select a different conductivity buffer during calibration (see page 21) · PGM key, used in programming
6
LO
TEMP. display for temperature in °C or °F and program #
7
THE REAR PANEL
RESET button and restart the entire operation.
Power connection Plug the 12VDC adapter (HI710005 or HI710006) into the DC socket (#6). Plug the adapter into the mains. Note: The instrument uses the following configuration.
1. DIN female socket for the conductivity probe (HI7686) 2. Temperature probe socket for use with the HI7669/2W 3. Reset button 4. Potentiometer for conductivity calibration (0.0µS) 5. RS 232C socket for serial communication. 6. DC Power Socket (for HI710005 or HI710006).
+
-
It is recommendable to use Hanna HI710005 or HI710006 voltage adapters (supplied with the meters) that use the proper polarity configuration. HI 9032 can also be used with other adapters. In this case, remember to check the correct polarity of your adapter before connecting it to the meter. Note: · Make sure the main line is protected by a fuse. · If you need to unplug the instrument, press the OFF key before disconnecting the meter from the mains.
Probes connection Connect the supplied conductivity probe HI7686 securely to the DIN socket provided (#1) by aligning the pins with the socket, pushing the plug in and tightening the threaded ring. For temperature measurements and automatic temperature compensation connect the temperature probe HI7669/2W to the °C socket (#2). RESET button The RESET button (#3) is used when the instrument displays erroneous messages due to strong electrical interference or when the instrument's power supply was disconnected before the meter was switched off. Press the
8
9
SPECIFICATIONS
OPERATIONAL GUIDE Make sure that the instrument has been calibrated before taking measurements (see calibration procedures on page 21 for conductivity, page 31 for NaCl and 35 for fertilizer). Connect the conductivity probe to the meter securely by aligning the pins with the socket on the back of the meter, pushing the plug in and tightening the threaded ring. Connect the temperature probe to the back of the meter for automatic temperature compensation or for temperature measurements. If possible, to minimize any EMC interferences, use plastic beakers for the solutions.
MEASURE COND. FERTILIZER NaCl TDS TEMP. RESOLUTION COND. FERTILIZER NaCl TDS TEMP. ACCURACY (@20°C/ 68°F) TEMP. TYPICAL EMC DEVIATION CALIBRATION
0.0-399.9 / 399-3999 µS/cm 3.99-39.99 / 39.9-99.9 mS/cm 0.00 to 39.99 g/l 0.0 to 130.0% 0.0-199.9 / 0-1999 ppm; 0.00 to 19.99 g/l 0.0 to 50.0°C / 32.0 to 122.0°F
0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm 0.01 g/l 0.1% 0.1 pmm; 1 ppm; 0.01 g/l 0.1°C / 0.1°F ±1% full scale each range ±0.5°C / ±1°F excluding probe error ±1% full scale all 4 ranges ±0.1°C / ±0.2°F Automatic single point at 84 µS, 1413 µS, 12.88 mS and 80.0 mS Automatic or manual from 0 to 50°C (32 to 122°F) with an adjustable ß from 0.0 to 3.0% per degree °C. Adjustable from 0.01 to 0.99 0 to 50°C (32 to 122°F); max. RH 95% 230 x 170 x 70mm (9.1 x 6.7 x 2.7") 1.5 kg (3.3 lbs)
TEMPERATURE COMPENSATION
TDS FACTOR ENVIRONMENT
DIMENSIONS
SHIPPING WEIGHT
TAKING CONDUCTIVITY, FERTILIZER, NaCl, TDS MEASUREMENTS With Automatic Temperature Compensation · Immerse the conductivity probe into the solution submerging the holes of the sleeve in the sample. Agitate the probe lightly in the solution to remove any air bubbles which may have been trapped inside the sleeve. Immerse the temperature probe as close as possible to the conductivity probe.
10
11
ON
·
Switch the instrument on by pressing the ON key. The meter automatically defaults to the conductivity mea- CONDUCTIVITY surement mode with 2.0% as temperature %TC coefficient (on the large LCD) and the temperature in the Celsius scale (on the TEMP. LCD). Select a different measurement mode by pressing the following keys:
mS
Conductivity measurement mode · By pressing the Cond. key, the display will show "CONDUCTIVITY" to confirm the measurement mode.
Cond.
CONDUCTIVITY
mS
7
%TC
·
·
Fert.
·
·
the Fert. key for fertilizer #1
8
The meter will automatically default the temperature coefficient % value to 2.0% (shown on the secondary display) every time it is turned on. Change the temperature coefficient to the value of the solution, by pressing the %TC key. The % value will start blinking on the secondary LCD.
%TC
CONDUCTIVITY
mS
Fert.
·
the Fert. key twice for fertilizer #2
8
.
%TC
NaCl
·
the NaCl key for NaCl of sea water saturation.
9
T.D.S.
·
the T.D.S. key for total dissolved solids
4
Cond.
·
the Cond. key to enter the conductivity measurement mode again.
7
Enter the desired value by pressing the numeric keys and then press the ENTER key to confirm the value. The stored value will be displayed and it will stop blinking. E.g. If you want to enter 2.5% as temperature coefficient value, proceed as follows: Press the %TC key Press 2 Press . Press 5
13
12
Press the ENTER key.
%TC
°F
%TC
ENTER
.
2
.
5
mS
If further measurements are desired, rinse the probe with a small amounts of the next sample and then proceed. Fertilizer measurement mode · By pressing the Fert. key the display will show "FERTILIZER" to confirm the measurement mode and the secondary display will show "1", i.e. Fertilizer #1.
Fert.
gm
·
CONDUCTIVITY
%TC
FERTILIZER
8
PGM
To clear the numeric entry, press the CLEAR key.
Clear
·
Select the second fertilizer ("2") by p ...