Detailed instructions for use are in the User's Guide.
UPW
HI98309
Ultra Pure Water Tester
OPERATION: · Remove the protective cap. · Rinse the tester with distilled water, up to the maximum immersion level. · Turn the UPW on. · Immerse it in the solution to be tested without exceeding the maximum immersion level. · Stir gently and wait for the reading to stabilize. · The reading on display is directly expressed in µS/cm unit. · To translate the reading into MegaOhm, refer to the following table. Distilled/Deionized Water Resistance (MegaOhm) 1 2 5 10 15 20 50 Equivalent UPW read-out (µS/cm) 1.000 0.500 0.200 0.100 0.067 0.050 0.020
display shows "1.500".
BEPS AND BATTERY REPLACEMENT: UPW is supplied with BEPS (Battery Error Preventing System) which avoids any erroneous reading due to low batteries level, by switching the meter off. To replace the batteries, slide off the battery compartment cover and replace all four 1.5V batteries while paying attention to their polarity.
Batteries should only be replaced in a nonhazardous area using the battery type specified in this instruction sheet.
ONE YEAR WARRANTY AND SERVICEABLE:
Covered by one year warranty against defects in workmanship and materials, this tester is now completely serviceable. Contact your dealer for further information. The UPW is in compliance with the CE directives.
CALIBRATION: All UPW testers are factory-calibrated with freshly prepared low-conductivity solutions. Infact , due to their high instability, calibration solutions in the UPW range (below 2 µS/cm) can not be supplied with the meter. However, if your meter needs to be recalibrated, use a 1.500 µS/cm solution prepared by a qualified laboratory and keep it airtight. · Rinse the electrode with some of the calibration solution.
Manufacturers since 1978
· Allow the reading to stabilize and with a small screwdriver turn the calibration trimmer until the
Visit our Internet Home Page: http://www.hannainst.com
ELECTRODE REPLACEMENT:
UPW
HI98309
Ultra Pure Water Tester
SPECIFICATIONS: RANGE 0.000 1.999 µS/cm RESOLUTION 0.001 µS/cm ACCURACY (@20°C/68°F) ±2% f.s. TYPICAL EMC DEVIATION ±0.5% f.s. ENVIRONMENT 0 to 50 C (32 to 122 F); 95% RH not condensing BATTERY TYPE 4 x 1.5V alkaline (included) BATTERY LIFE approx. 120 hours of continuous use DIMENSIONS 175 x 41 x 23 mm (7.9 x 1.8 x 1") WEIGHT 78 g (2.7 oz.)
Internal circuit:
The electrode is built-in inside the plastic casing and can be easily replaced in the following way: · Slide off the battery cover. · Remove the screw on the back of the UPW located below the battery compartment and slide off the whole internal part of the tester.
· Fasten the screw on the back, below the battery compartment.
ACCESSORIES: HI 73309 HI 731326 Spare electrode Calibration screwdriver (20 pcs) · Replace the built-in electrode with a new HI73309. · Before reinserting the internal circuit into the new plastic casing, make sure the O-rings (2) of the metal tube (3) are properly seated (see drawing). To avoid bending the connections to the conductivity sensors (1), do not apply excessive force while inserting the circuit. In case of any problem, reslide and reinsert it.
· Reinsert the battery compartment cover.
IS98309-R4