Detailed instructions for use are in the User's Guide.
Instruction Manual
pH 300 pH 301 pH 302
Microprocessor GLP Bench-top pH/mV/ISE/°C Meters
Rel mV ppm pH
DATE TIME
CFM MEM WRONG
°C mV ppm pH INTV
CAL LOG
1
BUF 2
3
m s
ON
OFF
PAPER
PRINT
pH302
pH METER
RANGE
CAL
CFM /
TIME
PGM
ENTER /
http://www.hannainst.com
These Instruments are in Compliance with the CE Directives
1
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with CSA, UL and (EN 50081-1 and EN 50082-1) directives.
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer. Each pH meter comes supplied complete with: · HI 1131B glass-body combination pH electrode · HI 7669/2W temperature probe · HI 3131B glass-body platinum ORP electrode · pH 4.01 & 7.01 buffer solutions (20 mL each) · 12VDC adapter (HI710005 or HI710006) · 5 paper rolls (pH 302 only) The KIT version includes also: · HI76405 electrode holder · HI 92000 Windows® compatible application software · HI 920010 PC connection cable · Storage, cleaning and electolyte solutions Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packing with the supplied accessories. GENERAL DESCRIPTION Hanna Instruments pH300, pH301 and pH 302 are professional bench-top pH meters for pH, ORP (Oxidation Reduction Potential), ISE and temperature measurements. Their built-in microprocessor ensures accurate and userfriendly operation. pH measurements are automatically compensated for the temperature effect (ATC). These meters come equipped with a large, easy-to-read LCD which shows the pH or mV or ION (for pH 301 only) with temperature simultaneously. The display has graphic symbols to make the calibration procedure easy to follow. Calibration is not allowed until the electrode has stabilized to ensure accurate measurements. Through the RS 232 serial port you can transfer sample measurements to a PC.
3
TABLE OF CONTENTS
Preliminary Examination ................................... 1 General Description .......................................... 1 Functional Description pH300 .......................... 2 Specifications pH300 ........................................ 4 Functional Description pH301 .......................... 5 Specifications pH301 ........................................ 7 Functional Description pH302 .......................... 8 Specifications pH302 ...................................... 10 The Rear Panel ................................................ 11 LCD Functional Description ............................. 11 Operational Guide ........................................... 12 pH Calibration ................................................. 29 pH Values at Various Temperature ................. 39 ION Calibration (for pH301 only) ..................... 40 mv Calibration ................................................. 44 Temperature Calibration ................................. 50 Logging with pH301 ........................................ 55 Programming Functions with pH302 .............. 58 Interface with PC ............................................. 82 Troubleshooting Guide ................................... 83 Electrode Conditioning and Maintenance ...... 84 Temperature-Resistance Correlation for HANNA pH Sensitive Glass ....................... 88 Additional Information about Interface with PC ............................................. 90 Accessories ..................................................... 99 Electrode Application Reference Guide ........ 107 Warranty ......................................................... 108 CE Declaration of Conformity ...................... 109
ISO 9000 Certified Company since 1992
2
FUNCTIONAL DESCRIPTION pH 300
THE FRONT PANEL
Keyboard on the Left · COMM to display the current baud rate for RS232. Keyboard on the Right · !°C and " °C to manually set temperature (when the temperature probe is not connected) and to select the calibration buffer, the baud rate, the command prefix. · CAL to enter and exit the calibration mode. After the pH calibration, to enter the calibration date, use this key to toggle between month, day and year. · CAL DATA to display the date of last pH calibration. This also displays the offset and the slope values expressed in mV. · CFM to confirm calibration values and to confirm baud rate and command prefix during communication setting mode. · MR to recall the stored value. · MS to hold and memorize the measured value. The last reading will be frozen on the display until the key is released. At the same time this reading is stored in the memory. · RANGE to select the measurement range (pH or mV).
Rel mV ppm pH
DATE TIME
CFM MEM WRONG
°C mV ppm pH INTV
CAL LOG
1
BUF 2
3
m s
ON
OFF
COMM
RANGE
CAL DATA
pH300
pH METER
CAL
°C
MS
CFM
°C
MR
Rel mV ppm pH
DATE TIME
CFM MEM WRONG
°C mV ppm pH INTV
CAL LOG
1
BUF 2
3
m s
ON
OFF
COMM
RANGE
CAL DATA
pH300
pH METER
CAL
°C
MS
CFM
°C
MR
· ON to turn the meter on · OFF to turn the meter off · Power LED to confirm the meter is on
4
5
SPECIFICATIONS pH 300
FUNCTIONAL DESCRIPTION pH 301
THE FRONT PANEL
Range pH ISE ORP °C pH ISE ORP °C 0.00 to 14.00 ±399.9 mV ±1999 mV -9.9 to 120.0 0.01 0.1 mV 1 mV 0.1 ±0.01 ±0.2 mV ±1mV ±0.5 ±0.01 ±0.6 mV ±1 mV ±0.5 Automatic 1 or 2 points with 3 memorized buffers (4.01, 7.01, 10.01) Max ±60mV or ±1pH 70-108% Automatic 1 or 2 points with 2 memorized values (0 and 50°C) Automatic or manual from -9.9 to 120°C (14 to 248°F) HI 1131B glass body combination refillable pH electrode (included) HI 7669/2W with 1 m (3.3') cable (included) 1012 RS 232 (optoisolated) 12 VDC (CSA, UL & CE approved) 0 to 50°C (32 to 122°F); 95% RH 280x200x75mm (11x7.9x3") 1.3 kg (2.9 lb.)
ON
CAL LOG
Rel mV ppm pH DATE TIME
Rel mV ppm pH
DATE TIME
CFM MEM WRONG
°C mV ppm pH INTV
Resolution
CAL LOG
1
BUF 2
3
m s
ON
OFF
Accuracy pH (@20°C/68°F) ISE ORP °C Typical EMC pH Deviation ISE ORP °C pH Calibration
LOG
CAL DATA
pH301
pH / ION METER
pH
ION
°C
0.00/0.000
mV
°C
pH Offset pH Slope Temperature Calibration Temperature Compensation Electrode Temperature probe Input impedance Computer Interface Power supply Environment Dimensions Shipping Weight
CFM MEM WRONG
°C mV ppm pH INTV
1
BUF 2
3
m s
OFF
LOG
CAL DATA
pH301
pH / ION METER
pH
ION
°C
CAL
DATE
COMM 0.00/0.000 mV °C
CFM
TIME
RANGE
· ON to turn the meter on · OFF to turn the meter off · Power LED to confirm the meter is on
7
6
Keyboard on the Left · CAL to start the calibration mode for pH, mV and ION. To set or quit the date or time setting mode. Used also during logging (to start logging press pH, mV or ION plus LOG plus CAL key; to stop logging press LOG plus CAL key). · CFM to confirm calibration values. · COMM to display the current baud rate and command prefix for RS232. · DATE to display the date. · RANGE to display the calibration buffer or the temperature on the secondary LCD during calibration. During setting of the date, to toggle between month, day and year. During setting of the time, to toggle between minutes, seconds and logging interval. · TIME to display the time. Keyboard on the Right · !°C and " °C to manually set temperature (when the temperature probe is not connected) and to select the calibration buffer, the date and time, the baud rate, the command prefix. · CAL DATA to display the date of last pH calibration. This also displays the offset and the slope 1 or slope 2 values expressed in mV. · ION to select the ION measurement range. · mV to select the mV measurement range. · pH to select the pH measurement range. · .00/.000 to select two or three decimal points for pH measurements.
SPECIFICATIONS pH 301
Range pH mV ppm °C -1.999 to 19.999 ±1999.9 0.001 to 19999 -9.9 to 120.0 0.01 or 0.001 0.1 0.001 (from 0.001 to 9.999) 0.01 (from 10.00 to 99.99) 0.1 (from 100.0 to 999.9) 1 (from 1000 to 19999) 0.1 ±0.002 ±0.1 (±799.9) / ±0.2 (±1999.9) ±0.5% ±0.5 ±0.01 ±0.3 ±1% ±0.5 Automatic 1, 2 or 3 points with 6 memorized buffers (1.68, 4.01, 6.86, 7.01, 9.18, 10.01) Max ±60mV or ±1pH 70-108% Automatic 1 or 2 points with 5 memorized buffers (0.1, 1, 10, 100, 1000) Max ±1999.9mV From ±20mV to ±100 mV [mV/log(0.001xppm)] Automatic 1 or 2 points with 2 memorized values (0 and 50°C) Automatic or manual (for pH only) from -9.9 to 120°C (14 to 248°F) HI 1131B glass body combination refillable pH electrode (included) HI 7669/2W with 1 m (3.3') cable (included) 1012
Resolution pH mV ppm
°C Accuracy pH (@20°C/68°F)m V ppm °C Typical EMC pH Deviation mV ppm °C pH Calibration
pH Offset pH Slope ppm Calibration
ppm Offset ppm Slope Temperature Calibration Temperature Compensation Electrode Temperature probe Input impedance Logging Interval
1, 15, 30 seconds or 1, 5, 30, 60, 120, 180 minutes PC Interface RS 232 (optoisolated) Power supply 12 VDC (CSA, UL & CE approved) Environment 0 to 50°C (32 to 122°F); 95% RH Dimensions 280x200x75mm (11x7.9x3") Shipping Weight 1.3 kg (2.9 lb.)
8
9
FUNCTIONAL DESCRIPTION pH 302
THE FRONT PANEL
Keyboard on the Left · 0 to 9 for numeric and decimal data input. · CLEAR to cancel wrong data entry during programming. Keyboard on the Right · CAL to perform the pH, mV or temperature calibration. · CFM/! to confirm pH/mV/temperature ! calibration values or to manually set temperature (when the temperature probe is not connected). · ENTER/"°C to terminate the numeric data " entry or to manually set temperature (when the temperature probe is not connected) or to select the buffer value during pH calibration. · PAPER to pull the paper. · PGM to select a program. · PRINT to obtain a printout (prints time, date, sample number and pH or mV values). · RANGE to select the measurement range (pH or mV) or to display the calibration buffer or the temperature on the secondary LCD during calibration. · TIME to display the time and the date.
Rel mV ppm pH
DATE TIME
CFM MEM WRONG
°C mV ppm pH INTV
CAL LOG
1
BUF 2
3
m s
ON
OFF
PAPER
PRINT
pH302
pH METER
RANGE
CAL
CFM /
TIME
PGM
ENTER /
Rel mV ppm pH
DATE TIME
CFM MEM WRONG
°C mV ppm pH INTV
CAL LOG
1
BUF 2
3
m s
ON
OFF
7
8
9
PAPER
PRINT
pH302
pH METER
4
5
6 RANGE CAL CFM /
1
2
3 TIME PGM ENTER /
0
.
CLEAR
· ON to turn the meter on · OFF to turn the meter off · Power LED to confirm the meter is on
10 11
SPECIFICATIONS pH 302
THE REAR PANEL
Range
pH ISE ORP °C pH ISE ORP °C
0.00 to 14.00 ±399.9 mV ±1999 mV -9.9 to 120.0 0.01 0.1 mV 1 mV 0.1 ±0.01 ±0.2 mV ±1mV ±0.5 ±0.01 ±0.3 mV ±1 mV ±0.5 Automatic 1, or 2 points with 3 memorized buffers (4.01,7.01,10.01) Automatic 1 or 2 points with 2 memorized values (0 and 50°C) Automatic or manual from -9.9 to 120°C (14 to 248°F) HI 1131B glass body combination refillable pH electrode (included) HI 7669/2W with 1 m (3.3') cable (included) 1012 Dot matrix, 44mm wide paper English, German, French, Spanish, Italian & Swedish RS 232 (optoisolated) 12 VDC (CSA, UL & CE approved) 0 to 50°C (32 to 122°F); 95% RH 280x200x75mm (11x7.9x3") 1.3 kg (2.9 lb.)
REFERENCE MEASUREMENT TEMPERATURE ELECTRODE ELECTRODE PROBE
RESET
12VDC
RS 232
ooooo oooo
Resolution
1
2
3
4
5
6
Accuracy pH (@20°C/68°F) ISE ORP °C Typical EMC pH Deviation ISE ORP °C pH Calibration
1. Reference Electrode Connector (Ø 4 mm) 2. BNC socket for measurement electrode 3. Socket for Temperature Probe 4. Reset Button 5. DC Power Socket (for HI710005 or HI710006) 6. RS 232 Connector
Temperature Calibration Temperature Compensation (for pH only) Electrode Temperature probe Input impedance Printer Languages PC Interface Power supply Environment Dimensions Shipping Weight
LCD FUNCTIONAL DESCRIPTION
Primary Display
Rel mV ppm pH DATE TIME
CFM MEM WRONG
°C mV ppm pH INTV
CAL LOG
1
BUF 2
3
m s
Secondary Display
12 13
OPERATIONAL GUIDE
ELECTRODE AND PROBE CONNECTIONS For combination pH or ISE or ORP electrodes (with internal reference) plug the electrode's BNC to the socket provided (see page 11). For electrode with separate reference, connect the measuring electrode's BNC to the BNC socket and the reference electrode's plug to the socket provided (see page 11). For temperature measurements and automatic temperature compensation connect the temperature probe to the °C socket (see page 11). If the temperature probe is not connected, when turned on, the meter will display a default temperature of 25°C or the last recorded temperature (if th ...