Detailed instructions for use are in the User's Guide.
Blood Pressure Monitor
Instruction Manual
Model Number 1092
ID #06/00
Requires 2 "AAA" Alkaline Batteries
2
YEAR LIMITED W ARRANTY
TABLE OF CONTENTS
Identification of Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Battery Installation/Replacement . . . . . . . . . . . . . . . . . . . . . . . . 4 Attaching the Wrist Cuff and Posture During . . . . . . . . . . . . . . . 4 Measurement Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Subsequent Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Automatic Power-off Function . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Memory Recall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Memory Clear Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Description of Display Marks . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Identification of Parts
ts
3
Battery Installation/Replacement
· To use the unit, you must install batteries. · Open battery cover as illustrated. Install or replace two new batteries, making sure the polarities (+ and -) on the battery match the polarities on the case. · Replace battery cover. · If the Low Battery indicator " " appears on the display, both batteries must be replaced. · "AAA" alkaline batteries or equivalent are suitable for this blood pressure meter.
Attaching the Wrist Cuff and Posture During Measurement
· The blood pressure is measured at the wrist. Therefore, rolling up the sleeve is no longer necessary. · Wrap the wrist cuff around your left wrist with the display panel on the inside of the wrist. · Pull the wrist cuff to fit wrist size and fasten velcro. Wrist circumference: 5.4-7.8 in.
Rest left elbow on a table so that the cuff is at heart level. Relax the lower arm across the table and remain still. Perform measurements in either a seated or reclining posture with the left palm turned upward.
1/2 inch
4
Measurement
NOTE: Unit measures on inflation. · To start the measurement Press the "Power/Start" switch.
· The heart symbol will flash and the buzzer will sound for 1 second. When the unit completes the preparations for measurement, the buzzer will sound three more times. If batteries are weak, "h " mark will not appear.
· Pressurization and measuring begin.
· When a pulse is detected, the pressure value moves to the lower side of the display. At this time the "h" mark flashes and buzzer beeps.
· Once the blood-pressure measurement has stabilized, pressurization is discontinued and the air in the cuff is automatically let out. · The buzzer sounds for about 2 seconds, and the high and low blood pressure are respectively dis-
5
played one above the other. · Unless an error is generated, the measurement results are automatically stored in memory (Up to 7 sets of results can be stored ). When there are already 7 sets of results in memory, the oldest one is automatically erased and overwritten with the newest.
· Blood pressure values and pulse rate are alternately displayed. The unit will continue to display the measurement until power is turned off.
· Turn Power off Press the "Power/Start" switch to turn off. NOTE: To discontinue a measurement for any reason, press the "Power/Start" switch. This causes the "wrist cuff" to rapidly deflate.
Subsequent Measurement
When a measurement needs to be made again, the time between 2 measurements must be at least 10 minutes.
Automatic Power OFF Function
If the unit is left on after measurement, an "Automatic PowerOFF Function" turns the power off after approximately 3 minutes. To continue with another measurement press the "Power/Start" switch to turn the unit on.
6
Memory Recall
· You can recall stored data by pressing the "Memory Recall" switch at any time, EXCEPT while taking a measurement. · Up to 7 sets of measurement results can be stored in memory. Memory cell No. 1 is always used to store the newest set of results.
· When pressing the "Memory Recall" switch, the memory cell number is displayed. The stored measurement results (systolic/diastolic blood pressure) are then displayed.
· If the "Memory Recall" switch is pressed again, the prior measurements are displayed in the order they were taken.
Memory Clear Function
While the memory data to be cleared is indicated and the "Memory recall" switch is pressed for 3 seconds, the buzzer will beep and the memory data is cleared.
7
Description of Display Marks
DISPLAY MARK CONDITION/CAUSE CORRECTIVE ACTION
The cuff was inflated to more than 280 mmHg.
Exhaust the air from the cuff.
Pressurization is not possible. The wrist cuff was not wrapped properly.
After rewrapping the cuff, repeat measurement. After rewrapping the cuff, conduct the measurement without moving. Repeat the measurement without moving the arm.
Pulse could not be measured accurately.
The arm moved during measurement.
There is a problem with the unit.
Return to Lumiscope Service Department. Repeat the measurement without moving the arm.
Pulse could not be measured.
Batteries are weak.
Replace the batteries.
There is a problem with the device.
Return to Lumiscope service Department.
8
Specifications
Model No. Measuring Principle Measurement Localization Cuff Wrist Circumference Range Measurement Range Accuracy LCD Indication 1092 Oscillometric Method Left Wrist Soft Cuff 5.4 to 7.8 in. 0 to 280 mmHg 40 to 180 Pulses/min. ±3 mmHg ±5% of reading
Pressure Pulse Pressure Pulse Pressure Pulse Rate
(13.5 to 19.5cm)
Systolic blood pressure: 3 digits Diastolic blood pressure: 3 digits Pulse rate: 3 digits
Alternatively
Marks LCD Switch Inflation Exhaust Rated Voltage Power Source Power Consumption Battery Life Automatic Power-Off Memory Function Dimension Weight Operating Environment Storage Environment
CAUTION:
View Size Digit Size
Temperature Humidity Temperature Humidity
Measuring h Battery 1.44(H) x .904(W)in (36(H) x 22.6(W)mm) .44(H) x .22(W)in (11(H) x 5.5(W)mm) Power/Start switch, Memory recall switch Automatic Inflation by Internal Pump Electromagnetic Valve 3 VDC 2 "AAA" Size Batteries 1.0W Approx. 400 times (180mmHg, 1time/day, 22°C) Approx. 3 min. Seven Memories Function 2.8(H) x 2.68(W) x 1.16(T)in (70(H) x 67(W) x 29(T)mm) Approx. 3.5 oz. (100g) 50 to 104°F (10 to 40°C) 30 to 85%RH -4 to 140°F (-20 to 60°C) 10 to 95%RH
1. Measurement of blood pressure may not be possible for someone with a weak pulse or cardiac arrhythmia. 2. This product is not intended to be used in the diagnosis or treatment of disease. 3. If you have high blood pressure and/or other cardiovascular conditions or diseases, this device should be used in consultation with your doctor. 4. Consult your doctor if you have questions about your blood pressure or pulse. 5. Specifications are subject to change without notice due to improvements in performance.
9
...