Detailed instructions for use are in the User's Guide. HMI38 HUMIDITY DATA PROCESSOR AND HMP35/36/37E PROBES Operating Manual
U063en-2.1 June 1998 © Vaisala 1998
© Vaisala Oyj 1998 No part of this document may be reproduced in any form or by any means, electronic or mechanical (including photocopying), nor may its contents be communicated to a third party without a prior written notice of the copyright holder. The instruction manuals may be changed without prior notice.
Yleisjäljennös 06/1998
U063en-2.1
HMI38 Operating Manual
Contents
1. 2. 3. 4. PRODUCT DESCRIPTION.......................................................................................... 1 TO BE NOTED WHEN MEASURING HUMIDITY ........................................................ 2 TAKING THE HMI38 INTO USE.................................................................................. 3 COMMISSIONING....................................................................................................... 4 4.1. 4.2. 4.3. 4.4. 5. Initializing the probe locations...................................................................... 4 Connecting the analogue outputs................................................................. 4 Connecting the RS 232C or RS 485 serial bus ............................................. 5 Summary of the factory settings ................................................................... 6
COMMANDS ............................................................................................................... 7 5.1. Display/keypad commands............................................................................ 7 5.1.1. Display mode .................................................................................... 7 5.1.2. Command mode ............................................................................... 7 5.1.3. Selecting and scaling the analogue outputs ...................................... 8 5.1.4. Setting the serial line parameters ...................................................... 9 5.1.5. Setting the output mode .................................................................. 11 5.1.6. Others ............................................................................................ 12 Serial commands ......................................................................................... 16 5.2.1. Analogue output commands............................................................ 16 5.2.2. Output via the serial bus ................................................................. 18 5.2.3. Operating the HMI38 via the serial bus ........................................... 20
5.2.
6.
DATA LOGGING....................................................................................................... 22 6.1. Logging 6.1.1. 6.1.2. 6.1.3. 6.1.4. using display/keypad commands ................................................. 23 Start/stop data logging in the CONTINUOUS mode ............................. 23 Start/stop data logging in the CATCH mode ...................................... 24 Outputting the logged data to the serial line .................................... 25 Outputting the logged data to the display ........................................ 26
6.2.
Logging using serial line commands (CONTINUOUS mode) ......................... 27 6.2.1. Data logging ................................................................................... 27
7.
CALIBRATION .......................................................................................................... 30 7.1. 7.2. Humidity calibration..................................................................................... 30 7.1.1. Calibration procedure...................................................................... 30 Temperature calibration .............................................................................. 31 7.2.1. With display/keypad commands: ..................................................... 31 7.2.2. With serial line commands .............................................................. 32 Calibrating the analogue outputs................................................................ 32
7.3. 8.
USING THE HMI38 AS A FIELD CALIBRATOR........................................................ 33 8.1. HMD/W60/70 and HMP140 series transmitters ........................................... 33 8.1.1. Calibrating transmitters with voltage outputs ................................... 34 8.1.2. Calibrating transmitters with current outputs ................................... 36 HMP230 series transmitters ........................................................................ 37 8.2.1. One point calibration ....................................................................... 38 i
8.2.
HMI38 Operating Manual
U063en-2.1
8.2.2. 8.3. 9.
Two point calibration ....................................................................... 39
HMD/W 20/30 and HMP130Y series transmitters ........................................ 40
MAINTENANCE ........................................................................................................ 42 9.1. 9.2. 9.3. Reference measurements ............................................................................ 42 Self-diagnostics ........................................................................................... 42 Adjusting the contrast of the display.......................................................... 43
10. TECHNICAL DATA ................................................................................................... 43 10.1. 10.2. 10.3. 10.4. 10.5. 10.6. 10.7. 10.8. 10.9. Relative humidity ......................................................................................... 43 Temperature ................................................................................................. 43 Calculated quantities ................................................................................... 43 Data logging ................................................................................................. 44 Analogue outputs......................................................................................... 44 Serial interface ............................................................................................. 44 Electronics ................................................................................................... 44 Mechanics .................................................................................................... 45 Electromagnetic compatibility .................................................................... 46 10.9.1. Emissions ....................................................................................... 46 10.9.2. Immunity......................................................................................... 46
11. ACCESSORIES AND SPARE PARTS ....................................................................... 46
APPENDIX 1: APPENDIX 2: APPENDIX 3: APPENDIX 4: APPENDIX 5:
SERIAL COMMANDS .......................................................................... 47 ERROR MESSAGES ............................................................................ 69 HMP35E PROBE.................................................................................. 75 HMP36E PROBE.................................................................................. 81 HMP37E PROBE.................................................................................. 87
WARRANTY ..................................................................................................................... 93
This manual is valid for programme version 2.01.
ii
U063en-2.1
HMI38 Operating Manual
1.
PRODUCT DESCRIPTION
The microprocessor based HMI38 Humidity Data Processor is a light, portable instrument with an optional carrying case and built-in rechargeable battery which make it suitable for spot checking measurements or for use as a transfer standard. The HMI38 is used together with Vaisala's HMP35E, HMP36E or HMP37E humidity and temperature probes to measure relative humidity and temperature quickly and accurately. Two probes can be connected to each HMI38. The HMI38 also calculates dewpoint, mixing ratio, absolute humidity and wet bulb temperature from the relative humidity and temperature it has measured. The HMI38 has two analogue outputs (0...1V or 0...5 V) which can be selected and scaled freely. The HMI38 also has an RS 232C and an RS 485 interface. The HMI38 Humidity Data Processor is used through the menu-based display. The user can choose between metric and non-metric units. Output modes are flexible: measured or calculated quantities can be displayed or they can be transferred to a computer or peripheral device using either RS 232C or RS 485 serial bus. Because the HMI38 units are addressable, there can be several units on the same RS 485 line. The HMI38 can be used as a one point field calibrator for HMP140 series and HMD/W60/70 transmitters with standard or customized settings, as well as for calibrating their predecessors HMD/W20/30 and HMP130Y series. It can also be used as a one or two point calibrator for HMP230 series transmitters. The transmitters are calibrated using an appropriate reference cable between the HMI38 and the calibrated transmitter. The HMI38 can also be used as a data logger. In continuous data logging, the user selects the interval and the starting time of logging; the HMI38 can store up to 254 readings from both probes in CONTINUOUS mode. In addition, the HMI38 can also be used for manual data logging. The CATCH mode enables the user to store readings manually at any given moment. In this mode, the HMI38 can store 127 readings from both probes. The HMI38 stores the measured relative humidity and temperature values. When the stored data is transferred to a computer or a peripheral through the serial bus, it calculates the other four quantities from them. The HMI38 is powered with DC voltage, but with an AC adapter it can be powered from the mains. HMI38 has a built-in battery and a charger.
1
HMI38 Operating Manual
U063en-2.1
2.
TO BE NOTED WHEN MEASURING HUMIDITY
It is essential in the measurement of humidity and especially in calibration that temperature equilibrium is reached. Even a small difference in temperature between the measured object and the sensor causes an error. If the temperature is +20 °C (+68 °F) and the relative humidity 50 %RH, a difference of ±1 °C between the measured object and the sensor causes an error of ±3 %RH. When the humidity is 90 %RH, the corresponding error is ±5.4 %RH. The error is at its greatest when the sensor is colder or warmer than the surroundings and the humidity is high. A temperature difference of a few degrees can cause water to condense on the sensor surface. In an unventilated space evaporation may take hours; good ventilation accelerates evaporation. The HUMICAP® sensor starts to function normally as soon as the water has evaporated. If the condensed water is contaminated, the life span of the sensor may shorten and calibration may change.
10 9 8 7
dRH (%RH)
6 5 4 3 2 1 0 -40 -20 0 20 40 60 80 100
Temperature (°C)
Figure 2.1 Measurement error at 100 %RH when the difference between the ambient and sensor temperature is 1 °C
2
U063en-2.1
HMI38 Operating Manual
3.
TAKING THE HMI38 INTO USE
Please, note that when you take the HMI38 into use for the first time, both probe locations have the factory setting for the probe type HMP35E. If this is not the case, the HMI38 displays error message(s) after power up (see appendix 3). Error messages can be cleared by pressing ENT button several times. Should this happen, you must initialize the probe locations. During initialization, the HMI38 gets information on which probe locations have probes connected to them. Initializing is described in Chapter 4.1. If you use the instrument as a data logger, you must enter the correct date and time; see Chapter 6. In order to enable the HMI38 to calculate the mixing ratio and wet bulb temperature accurately, enter the ambient air pressure through the keypad or serial bus. This setting is described in Chapter 5. The HMI38 has a rechargeable battery which has not been charged in the factory. Charge the battery overnight. You can use the instrument instantly when you connect it to the mains for charging. To maximize the battery lifetime, completely discharge and recharge the battery about once a month. You can discharge the battery by leaving the HMI38 on without an AC adapter until the low battery voltage switches the HMI38 automatically off.
3
HMI38 Operating Manual
U063en-2.1
4.
4.1.
COMMISSIONING
Initializing the probe locations You can use either keypad or serial bus commands for the initialization of the probe locations. If you use serial commands, see Appendix 1. If the probe type setting in the HMI38 memory is not correct, some error messages may appear on the display after power up (see Appendix 3). You can clear these messages by pressing ENT button one or several times. When the humidity and temperature readings appear on the display, press CL key to enter the command mode. The first display is the main menu:
Use arrow keys to select Init which starts to blink, and then press ENT. The following appears with the indication of PROBE 1 blinking (in this example, HMP35E):
Use arrow keys to select the correct type for PROBE 1 (HMP35E, HMP36E, HMP37E, HMP35BM, HMP36BM) and press ENT. If there is no probe in location 1, select none and press ENT. Use arrow keys to select also the type for PROBE 2 and press ENT. The HMI38 returns to the display mode showing the current readings. 4.2. Connecting the analogue outputs Figure 4.1 shows the connections to the analogue output connector. See Chapter 5.1.3 and Chapter 5.1.3.2 for selecting the quantities and for scaling the analogue outputs.
ANALOGUE OUTPUT
0-1V/0-5V
1 2 3 4 5 6 7 8 9
AGND CH1 OUT NC CH2 OUT AGND NC NC NC NC D9S
Figure 4.1
Cable connections to the analogue output connector
4
U063en-2.1
HMI38 Operating Manual
4.3.
Connecting the RS 232C or RS 485 serial bus Figure 4.2 shows the pin configuration of the serial line connector and the RS 232C serial line connection of the HMI38 to a peripheral device.
HMI38 SER I/O
DO NOT CONNECT RS 232 C RXD RS 232 C TXD NC GND NC NC RS 485 HI RS 485 LO D9P D9S
1 2 3 4 5 6 7 8 9
COMPUTER (IBM AT) SER I/O
2 3 4 5 6 7 8 9 1
RXD TXD DTR GND DSR RTS CTS RI DCD
DO NOT CONNECT RS 232 C RXD RS 232 C TXD NC GND NC NC RS 485 HI RS 485 LO
HMI38 SERIAL PRINTER SER I/O (RS 232 PANASONIC KXSER I/O 1
2 3 4 5 6 7 8 9
2 3 4 5 6 7 8 20 22
TXD RXD NC NC DSR GND DCD DTR NC D25P
D9P
Figure 4.2
Wiring between the HMI38 and a PC or a serial line printer using RS 232C serial line.
Connect the serial bus between the PC and the HMI38 and switch the PC on. Use a terminal emulation programme (e.g. Procomm Plus, Datastorm or Windows terminal) when using a PC. The following chapter lists the factory settings for data transfer.
NOTE
When changing the serial bus settings, switch the instrument off and on again or give the serial line command RESET to activate the new settings. Refer to the manuals of the PC and the terminal emulation programme when giving serial settings.
5
HMI38 Operating Manual
U063en-2.1
4.4.
Summary of the factory settings The factory settings of the HMI38 are the following: · probe initializations: -probe 1 -probe 2 auto power-off output units pressure analogue outputs: -channel 1 RH1: -channel 2 T1: serial line: -bauds, parity, data, stop -duplexity address serial mode echo output interval
HMP35E HMP35E off metric 1013.25 hPa (= mbar) 0...100 %RH 0...1 V -40...160 °C 0...1 V 4800, even, 7, 1 FDX 0 stop on 0s
· · · · ·
· · · · · ·
measurement integration time 16 filter off
6
U063en-2.1
HMI38 Operating Manual
5.
COMMANDS
The HMI38 incorporates a microprocessor which allows the configuration to be set according to the user's needs. This is done through commands, either utilizing the menus on the front panel display or giving commands through the serial interface. The commands can be used e.g. to choose the displayed quantities, to select and scale the analogue outputs, to calibrate the temperature channels as well as the analogue outputs and to set the serial interface.
5.1.
Display/keypad commands 5.1.1. Display mode In the display mode the HMI38 outputs measured and calculated quantities on the display. You can scroll the first line with button s and the second line with button t. After reset the instrument shows first the programme type and version and then goes on to the display mode.
You can scroll different quantities with arrow switches. Acknowledge each selection by pressing ENT; this way, they appear on the display even after power off. The display also shows error messages and alarms if they occur. 5.1.2. Command mode Press the CL key to enter the command mode. The first display is the main menu:
You can scroll the commands with the arrow keys. The currently active command blinks; select the desired command with the ENT key. When a menu is on the display, either the first command or the currently valid setting blinks. The CL key takes the HMI38 back to the display mode.
7
HMI38 Operating Manual
U063en-2.1
5.1.3. Selecting and scaling the analogue outputs 5.1.3.1.
·
Selecting the output Select Mode in the main menu and Analog outputs in the Mode menu:
·
Select Mode 1V / 5V. The current settings for channel 1 are displayed:
· ·
If the settings are correct, press ENT. If you need to change the settings, press CL:
- the lower limit starts blinking - acknowledge the lower limit with ENT or change it with the nu- the upper limit starts blinking - acknowledge the upper limit with ENT or change it with the numeric keys, then press ENT meric keys, then press ENT
·
When channel 1 has been set, the programme goes on to channel 2; the procedure is the same as with channel 1.
NOTE
The hardware has two basic voltage ranges: 0...1 V and 0...5 V. When scaling analogue outputs, please note that the relative accuracy of the output may become too low if only a very small part of either range is used to output a quantity with a wide measurement range. 5.1.3.2.
·
Selecting and scaling the analogue output quantities Select Mode in the main menu and Analog outputs in the Mode menu:
·
Select Scale. The quantity and scaling for channel 1 are displayed:
8
U063en-2.1
HMI38 Operating Manual
· ·
If the settings are correct, press ENT. If you need to change the settings, press CL:
- the quantity (RHn, Tn, Tdn, xn, an, Twn; n = 1 or 2) starts blink- the lower limit starts blinking - acknowledge the lower limit with ENT or give a new value using - the upper limit starts blinking - acknowledge the upper limit with ENT or give a new value using
the numeric keypad and press ENT the numeric keypad and press ENT ing; change it with the arrow keys and acknowledged with the ENT key
·
After channel 1 has been set the settings of channel 2 appear on the display; follow the procedure described above for channel 1.
5.1.4. Setting the serial line parameters 5.1.4.1.
·
Turning the serial interface echo ON/OFF Select More in the main menu, More in the More menu and Echo in the second More menu.
·
Use the arrow keys to select the right alternative and press ENT. Serial bus settings Select Seri in the main menu; the currently valid serial interface settings are displayed:
5.1.4.2.
·
· ·
If the settings are correct, press ENT; the programme returns to the display mode. If you need to change the settings, press CL:
9
HMI38 Operating Manual
U063en-2.1
·
Select the parameter to be changed with the arrow keys and ENT key.
- Selecting baud rate:
- Selecting parity:
- Selecting data bits:
- Selecting stop bits:
- Full duplex/half duplex:
NOTE
The serial bus settings become effective only after power off/on or the serial line command RESET. 5.1.4.3. Setting the instrument address
Address is used when more than one instrument is connected to one serial bus; this enables communication with one instrument at a time.
·
Select More in the Main menu and then Addr in the More menu; the following appears on the display:
· ·
Change the address by entering a new value and pressing ENT. Pressing ENT returns the programme to the More menu.
10
U063en-2.1
HMI38 Operating Manual
·
Pressing CL returns the programme to display mode and leaves the address unchanged. Selecting the output units Select More in the Main menu and Unit in the More menu:
5.1.4.4.
·
·
Use the arrow keys to select the right alternative and press ENT. RH T Td a x Tw metric %RH °C °C g/m3 g/kg °C non-metric %RH °F °F gr/ft3 gr/lb °F
5.1.5. Setting the output mode The output modes only affect output through the serial interface: the instrument accepts all display/keypad commands irrespective of which serial output mode it is in. The HMI38 has three serial output modes: RUN, STOP and POLL. In the RUN mode the instrument outputs measurements automatically through the serial interface to a PC or a peripheral. The only command that can be given through the serial interface is S (stop), which ends the RUN mode. In the STOP mode serial commands are given to the instrument. Measurements can be output only by giving command SEND. The POLL mode is used when more than one instrument is connected to the same serial bus; a single instrument can be addressed and communicated with. When the connection to one instrument is opened in the POLL mode, this instrument goes into STOP mode and can then accept commands normally. Closing the connection returns the instrument to POLL mode. In POLL mode the instrument outputs measurements only when requested (command SEND aa). If the user has forgotten the address of the instrument, it can be checked using display/keypad commands.
11
HMI38 Operating Manual
U063en-2.1
5.1.5.1.
·
Setting the serial interface operation mode Select Mode in the main menu; the following appears on the display:
·
Select Serial:
·
The currently valid setting blinks. Select the desired mode with the arrow keys and press ENT. After this the programme returns to the previous Menu. If you select the Run mode, the currently valid output interval appears on the display (with number and unit blinking):
·
If required, change the output interval setting as follows:
· · · · · ·
press CL the number starts blinking if you need to change the interval, enter the new interval; otherwise press ENT the unit (seconds, minutes or hours) starts blinking if necessary, change the unit with the arrow keys and acknowledge it with ENT after this the programme returns to the previous menu
5.1.6. Others 5.1.6.1. Setting the measurement integration time
Lengthening of the measurement integration time filters out any stray changes in the output: the instrument calculates the average of the number of measurement cycles defined by the user.
·
Select More in the main menu, More in the More menu and Mtim in the second More menu:
12
U063en-2.1
HMI38 Operating Manual
· ·
Pressing ENT returns the programme to the previous menu without changing the integration time. If you need to change the integration time, enter the new time (4...255) and press ENT.
5.1.6.2.
Setting the measurement integration filter
If the integration filter is set on, the output reading is an average of four last measurements.
·
Select More in the main menu, More in the More menu and Filt in the second More menu:
·
Choose the correct alternative (Filter on/off) with arrow keys and press ENT. Setting the pressure for mixing ratio and wet bulb temperature calculations
5.1.6.3.
The atmospheric pressure has an effect on mixing ratio and on wet bulb temperature. To achieve accurate calculations, the instrument needs to take the ambient pressure into account.
·
Select More in the main menu:
·
Select Pres:
· ·
Pressing ENT returns the programme to the previous menu without changing the pressure reading. If you need to change the pressure, enter the new value and press ENT.
13
HMI38 Operating Manual
U063en-2.1
5.1.6.4.
·
Setting the date Select More in the main menu; select Date in the More menu:
· ·
If the date is correct, acknowledge it by pressing ENT; this takes the programme back to the More menu. If you need to change the date, press CL. Enter the new date in the order the digits blink (year-month-day) and press ENT. Press ENT again to return to the previous menu. If there is no need to change the date, press CL. Setting the time Select More in the main menu; select Time in the More menu:
5.1.6.5.
·
· ·
If the time is correct, acknowledge it by pressing ENT; this takes the programme back to the More menu. If you need to change the time, press CL. Enter the new time in the order the numbers blink and press ENT. Press ENT again to return to the previous menu. If there is no need to change the time, press CL.
NOTE
As the HMI38 does not have a real-time clock with a backup battery, the date and time settings do not stay in the memory during power-off. 5.1.6.6. Setting the automatic shutdown
You can set the HMI38 to switch the power off automatically if no button is pressed for a certain time.
·
Select More in the main menu; select More in the first More menu and then again in the second More menu:
·
Press ENT and select the correct alternative (on/off) for the power-off mode with the arrow keys, then press ENT.
14
U063en-2.1
HMI38 Operating Manual
·
If you choose On, the current power-off time appears. If required, change the time setting as follows:
· · · · · ·
press CL the number starts blinking if you need to change the time, enter the new time; otherwise press ENT the unit (seconds, minutes or hours) starts blinking if required, change the unit with the arrow keys and acknowledge it with ENT after this the programme returns to Main menu
NOTE
Auto power-off is not operational during data logging and calibration procedures.
15
HMI38 Operating Manual
U063en-2.1
5.2.
Serial commands Appendix 1 contains more detailed descriptions of the serial commands. The following chapters include only the most commonly used command sequences. You can interrupt any serial command while giving it by pressing ESC. In the commands
If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual VAISALA HMP35 PROBES. VAISALA offer a product for which we do not have the user manual? Let us know what you are looking for: site Internet, histoire, actualité, filiales, site Internet, mode d'emploi, driver, avis des utilisateurs, meilleur prix des produits, forum d'assistance aux problèmes, annuaire des marques, annuaire des constructeurs, répertoire des marques, répertoire des constructeurs, site Internet de la marque, site Internet du constructeur Diplodocs allows you to download user manual VAISALA HMP35 PROBES, user guide VAISALA HMP35 PROBES, instructions VAISALA HMP35 PROBES, owner's manual VAISALA HMP35 PROBES, online manual VAISALA HMP35 PROBES.Access web reviews VAISALA HMP35 PROBES, , |
![]() |
Include the add-on to download manuals from your site, forum or blog | ![]() |
Frequently Asked Questions | ![]() |
Contact Diplodocs team | ![]() |
Last searches Last additions |
![]() |
Sitemap | ![]() |
|||
| Brands starting with A B C D E F G H I J K L M N O P Q R S T U V W X Y Z # | |||||||||||||
|
|
Copyright © 2005 - 2008 - Diplodocs -
All Rights Reserved. Designated trademarks and brands are the property of their respective owners. |